So Youre Considering Saline-Filled Breast Implant Surgery
What Gives the Breast Its Shape?
What Is A Saline-Filled Breast Implant?
What Are Important Factors for You to Consider When Deciding to Have Saline-Filled Implants?
Are You Eligible for Saline-Filled Breast Implants?
Who is not Eligible for Breast Implants?
What are Contraindications, Warnings, and Precautions for You to Consider?
What Types of Breast Implants Are Available from INAMED Aesthetics?
What are the Potential Breast Implant Complications?
Breast Augmentation Results from A95 Study
What Were the Complication Rates from the A95 Study?
What Were the Types of Additional Surgical Procedures Performed?
What Were the Reasons for Implant Removal?
What Were the Complication Rates After Implant Replacement?
Breast Augmentation Considerations
Special Considerations for Breast Augmentation
What Questions Should You Ask Your Surgeon About Breast Augmentation?
Breast Reconstruction Results from R95 Study
What Were The Complication Rates from The R95 Study?
What Were the Types of Additional Surgical Procedures Performed?
What Were the Reasons for Implant Removal?
What Were the Complication Rates After Implant Replacement?
Breast Reconstruction Considerations
Special Considerations for Breast Reconstruction
What Questions Should You Ask Your Surgeon About Breast Reconstruction?
If You Experience a Problem, Should You Report It?
What Are Other Sources of Additional Information?
So
Youre Considering
|
The purpose of this brochure is to assist you in making an informed decision about breast augmentation and breast reconstruction surgery. This educational brochure is set up to help you talk with your surgeon, as well as provide you with general information on breast implant surgery and give you specific details about McGhan® Saline-Filled Breast Implants. |
The breast consists of milk ducts and glands, surrounded by fatty tissue that
provides its shape and feel. Situated beneath the breast is the pectoralis major
muscle (chest muscle) of the chest wall. Factors such as pregnancy (when milk
glands are temporarily enlarged), rapid weight loss, and the effects of gravity
as you age combine to stretch the skin, which may cause the breast to droop
or sag.
A breast implant is a sac (implant shell) of silicone elastomer (rubber), which is surgically implanted under your tissues, and then filled with sterile saline, a salt water solution, through a valve.
IMPLANTS ARE TO BE USED FOR FEMALES FOR THE FOLLOWING INDICATIONS (PROCEDURES):
BREAST AUGMENTATION
This is done to increase the size and proportion of a womans breasts. A
woman must be at least 18 years old for breast augmentation.
BREAST RECONSTRUCTION
This procedure is done to restore a womans breast shape after a mastectomy or injury
that resulted in either partial or total loss of the breast(s), or to correct a birth
defect.
IMPLANTS ARE NOT TO BE USED FOR:
SURGICAL PRACTICES THAT ARE CONTRAINDICATED IN BREAST IMPLANTATION BECAUSE THEY MAY DAMAGE THE SHELL AND CAUSE DEFLATION/RUPTURE:
SAFETY AND EFFECTIVENESS HAS NOT BEEN ESTABLISHED IN PATIENTS WITH THE FOLLOWING CONDITIONS:
FURTHER CONSIDERATIONS:
Pre-implantation Mammography
You may wish to undergo a preoperative mammogram and another one 6 months to one year
after implantation to establish a baseline.
Interference with Mammography
The implant may interfere with finding breast cancer during mammography and also may make
it difficult to perform mammography. Therefore, it is essential that you tell your
mammography technologist that you have an implant before the procedure. The technologist
can use special techniques to minimize the possibility of rupture and to get the best
possible views of the breast tissue. Because the breast is squeezed during mammography, it
is possible for an implant to rupture during the procedure. More xray views are necessary
with these special techniques; therefore, women with breast implants will receive more
radiation. However, the benefit of the mammogram in finding cancer outweighs the risk of
the additional x-rays.
Distinguishing the implant from breast tissue during breast self-examination
You should perform a breast self-examination monthly on your implanted breast. In order to
do this effectively, you should ask your surgeon to help you distinguish the implant from
your breast tissue. Any new lumps or an abnormal finding on the mammogram should be
evaluated with a biopsy. If a biopsy is performed, care must be taken to avoid puncturing
the implant.
Long Term Effects
The long term safety and effectiveness of breast implants have not been studied; however,
INAMED is monitoring the long term (10 year) chance of implant rupture, reoperation,
implant removal, and capsular contracture (hardening of the tissues around the implant).
INAMED is also conducting mechanical testing to assess the long-term likelihood of implant
rupture. As new information becomes available, INAMED will issue an updated version of
this brochure.
Capsulotomy
You should be aware that closed capsulotomy, the practice of forcible squeezing or
pressing on the fibrous capsule around the implant to break the scar capsule is not
recommended as this may result in breakage of the implant.
Breast implants come in a variety of shapes, surface textures, and sizes. All currently available INAMED Aesthetics implants have a selfsealing (diaphragm) valve that is used for filling the device. Depending on the style, the filling valve may be located on the front (anterior) or the back (posterior) of the implant. Below is a description of INAMED Aesthetics breast implant styles. Be sure to familiarize yourself with the different features of breast implants and to discuss the most appropriate type(s) of implants for you with your surgeon.
Round Breast Implants: | |
Style 68: | Smooth shell surface, anterior filling valve, moderate projection |
Style 168: | BIOCELL® Textured shell surface, anterior filling valve, moderate projection |
Shaped Breast Implants: | |
Style 163: | BIOCELL® Textured shell surface, posterior filling valve, full height, full projection |
Style 363: | BIOCELL® Textured shell surface, anterior filling valve, low height, full projection |
Style 468: | BIOCELL® Textured shell surface, anterior filling valve, full height, moderate projection |
A = Width; B = Projection Round Breast Implant |
A = Width; B = Height; C= Projection Shaped Breast Implant |
Undergoing any surgical procedure may involve the risk of complications such as the effects of anesthesia, infection, swelling, redness, bleeding, and pain.
IN ADDITION, THERE ARE POTENTIAL COMPLICATIONS SPECIFIC TO BREAST IMPLANTS. THESE COMPLICATIONS INCLUDE:
Deflation/Rupture
Breast implants deflate when the saline solution leaks either through an unsealed or
damaged valve or through a break in the implant shell. Implant deflation can occur
immediately or slowly over a period of days and is noticed by loss of size or shape of
your breast. Some implants deflate (or rupture) in the first few months after being
implanted and some deflate after several years. Causes of deflation include damage
by surgical instruments during surgery, overfilling or underfilling of the implant with
saline solution, capsular contracture, closed capsulotomy, stresses such as trauma or
intense physical manipulation, excessive compression during mammographic imaging,
umbilical incision placement, and unknown/unexplained reasons. You should also be aware
that the breast implant may wear out over time and deflate/rupture.
Deflated implants require additional surgery to remove and to possibly replace the implant.
Capsular Contracture
The scar tissue or capsule that normally forms around the implant may tighten and squeeze
the implant and is called capsular contracture. Capsular contracture may be more common
following infection, hematoma, and seroma. It is also more common with subglandular
placement (behind the mammary gland and on top of the chest). Symptoms range from mild
firmness and mild discomfort to severe pain, distorted shape, palpability of the implant,
and/or movement of the implant.
Additional surgery is needed in cases where pain and/or firmness is severe. This surgery ranges from removal of the implant capsule tissue to removal and possibly replacement of the implant itself.
Capsular contracture may happen again after these additional surgeries.
Pain
Pain of varying intensity and duration may occur and persist following breast
implant surgery. In addition, improper size, placement, surgical technique, or capsular
contracture may result in pain associated with nerve entrapment or interference with
muscle motion. You should tell your surgeon about severe pain.
Additional Surgeries
You should should know that there is a high chance that you will need to have additional
surgery at some point to replace or remove the implant. Also, problems such as deflation,
capsular contracture, infection, shifting, and calcium deposits can require removal of the
implants. Many women decide to have the implants replaced, but some women do not. If you
choose not to, you may have cosmetically unacceptable dimpling and/or puckering of the
breast following removal of the implant.
Dissatisfaction with Cosmetic Results
Dissatisfying results such as wrinkling, asymmetry, implant displacement
(shifting), incorrect size, unanticipated shape, implant palpability, scar deformity,
hypertrophic (irregular, raised scar) scarring, and/or sloshing may occur. Careful
surgical planning and technique can minimize but not always prevent such results.
Infection
Infection can occur with any surgery. Most infections resulting from surgery appear within
a few days to weeks after the operation. However, infection is possible at any time after
surgery. Infections with an implant present are harder to treat than infections in normal
body tissues. If an infection does not respond to antibiotics, the implant may have to be
removed, and another implant may be placed after the infection is resolved.
In rare instances, Toxic Shock Syndrome has been noted in women after breast implant surgery, and it is a life-threatening condition. Symptoms include sudden fever, vomiting, diarrhea, fainting, dizziness, and/or sunburn-like rash. A surgeon should be seen immediately for diagnosis and treatment for this condition.
Hematoma/Seroma
Hematoma is a collection of blood inside a body cavity, and a seroma is a collection of
the watery portion of the blood (in this case, around the implant or around the incision).
Postoperative hematoma and seroma may contribute to infection and/or capsular contracture.
Swelling, pain, and bruising may result. If a hematoma occurs, it will usually be soon
after surgery, however this can also occur at any time after injury to the breast. While
the body absorbs small hematomas and seromas, large ones will require the placement of
surgical drains for proper healing. A small scar can result from surgical draining.
Implant deflation/rupture can occur from surgical draining if damage to the implant occurs
during the draining procedure.
Changes in Nipple and Breast Sensation
Feeling in the nipple and breast can increase or decrease after implant surgery. The range
of changes varies from intense sensitivity to no feeling in the nipple or breast following
surgery. Changes in feeling can be temporary or permanent and may affect your sexual
response or your ability to nurse a baby. (See the paragraph on breast-feeding below.)
Breast Feeding
At this time it is not known if a small amount of silicone may diffuse (pass through) from
the saline-filled breast implant silicone shell and may find its way into breast milk. If
this occurs, it is not known what effect it may have on the nursing infant. Although there
are no current methods for detecting silicone levels in breast milk, a study measuring
silicon (one component in silicone) levels did not indicate higher levels in breast milk
from women with siliconefilled gel implants when compared to women without implants.
With respect to the ability to successfully breast feed after breast implantation, one study reported up to 64% of women with implants who were unable to breast feed compared to 7% without implants. The periareolar incision site may significantly reduce the ability to successfully breast feed.
Calcium Deposits in the Tissue Around the Implant
Deposits of calcium can be seen on mammograms and can be mistaken for possible cancer,
resulting in additional surgery for biopsy and/or removal of the implant to distinguish
the calcium deposits from cancer.
Delayed Wound Healing
In some cases, the incision site takes longer to heal than normally.
Extrusion
Unstable or compromised tissue covering and/or interruption of wound healing may result in
extrusion, which is when the breast implant comes through the skin.
Necrosis
Necrosis is the formation of dead tissue around the implant. This may prevent wound
healing and require surgical correction and/or implant removal. Permanent scar deformity
may occur following necrosis. Factors associated with increased necrosis include
infection, use of steroids in the surgical pocket, smoking, chemotherapy/radiation, and
excessive heat or cold therapy.
Breast Tissue Atrophy/Chest Wall Deformity
The pressure of the breast implant may cause the breast tissue to thin and shrink. This
can occur while implants are still in place or following implant removal without
replacement.
In addition to these complications, there have been concerns with certain systemic diseases, of which you should be aware:
Connective Tissue Disease
Concern over the association of breast implants to the development of autoimmune or
connective tissue diseases, such as lupus, scleroderma, or rheumatoid arthritis, was
raised because of cases reported in the literature with small numbers of women with
implants. A review of several large epidemiological studies of women with and without
implants indicates that these diseases are no more common in women with implants than
those in women without implants. However, a lot of women with breast implants believe that
their implants caused a connective tissue disease.
Cancer
Published studies indicate that breast cancer is no more common in women with implants
than those without implants.
Second Generation Effects
There have been concerns raised regarding potential damaging effects on children born of
mothers with implants. A review of the published literature on this issue suggests that
the information is insufficient to draw definitive conclusions.
Although you will experience your own risks (complications) and benefits following breast implant surgery, this section describes the specific complications and benefits of McGhan® saline-filled breastimplants. INAMEDs studies indicate, for example, that most women can expect to experience at least one complication at some point through 5 years after implant surgery. The studies also indicate that the chance of additional surgery through 5 years is about 1 in 4 for augmentation patients and 1 in 2.5 for reconstruction patients. The information below provides more details about the complications and benefits you may experience.
INAMED conducted clinical studies testing of its saline-filled breast implants to determine the short-term and most common complications as well as benefits of their implants. These were assessed in the following studies:
The Large Simple Trial was designed to determine the 1-year rates of capsular contracture, infection, implant leakage/deflation, and implant replacement/removal. There were 2,333 patients enrolled for augmentation, 225 for reconstruction, and 317 for revision (replacement of existing implants). Of these enrolled patients, 62% returned for their 1-year follow-up visit.
The A95 and R95 Studies were designed as 5-year studies to assess all complications as well as patient satisfaction, body image, body esteem, and self concept. Patients were followed annually and data through 3 years (with partial 4 year data) were presented to FDA for PMA approval. After PMA approval, INAMED transitioned data collection to a post-approval study. The first phase of this postapproval study consisted of completion of the A95 and R95 Studies, with collection of all risk/benefit information through 5 years.
This section of the brochure includes original data presented to FDA for PMA approval along with the 5-year post-approval study data. The data presented through 5 years includes earlier data shown in the tables with new information added to it.
The A95 Study enrolled 901 augmentation patients, with 77% returning for their 3-year follow-up visit. Of those A95 patients available to be seen for their 5-year follow-up visit, 81% returned and were seen at 5 years after implant surgery. The R95 Study enrolled 237 reconstruction patients, with 71% returning for their 3-year follow-up visit. Of those R95 patients available to be seen for their 5-year followup visit, 80% returned and were seen at 5 years after implant surgery.
The table below shows the complication rates for augmentation, reconstruction, and revision patients through 1 year. The rates reflect the number of patients out of 100 who experienced the listed complication. For example, 7% or 7 out of 100 augmentation patients experienced capsular contracture at some time within 1 year after implantation. However, this does not mean that 7% of patients still have capsular contracture at 1 year.
Complication |
1-Year Complication Rate* |
||
Augmentation |
Reconstruction |
Revision |
|
Capsular Contracture III/IV |
7% |
13% |
12% |
Implant Removal |
6% |
14% |
8% |
Leakage/Deflation |
4% |
3% |
5% |
Infection |
2% |
6% |
3% |
*Data on 62% of the 2875 patients enrolled in the study
The 3-year and 5-year complication rates are shown from the most common 5-year rate to the least common 5-year rate in the table below. The rates reflect the number of augmentation patients out of 100 who experienced the listed complication at least once within the first 3 and 5 years after implantation. Some complications occurred more than once for some patients. The two most common complications experienced within the first 5 years of implantation were reoperation (26% or 26 patients out of 100) and breast pain (17% or 17 patients out of 100).
Complications |
N = 901 Patients |
|
3-Year** Complication Rate |
5-Year Complication Rate |
|
Additional Operation (Reoperation) | 21% | 26% |
Breast Pain* | 16% | 17% |
Wrinkling* | 11% | 14% |
Asymmetry* | 10% | 12% |
Implant Palpability/Visibility* | 9% | 12% |
Implant Replacement/Removal for Any Reason | 8% | 12% |
Capsular Contracture | 9% | 11% |
Intense Nipple Sensation* | 9% | 10% |
Loss of Nipple Sensation* | 8% | 10% |
Implant Malposition* | 8% | 9% |
Intense Skin Sensation* | 7% | 8% |
Scarring Complications | 6% | 7% |
Leakage/Deflation | 5% | 7% |
Irritation/Inflammation* | 3% | 3% |
Seroma | 3% | 3% |
Hematoma | 2% | 2% |
Skin Rash | 2% | 2% |
Capsule Calcification* | 1% | 2% |
Delayed Wound Healing* | 1% | 1% |
Infection | <1% | 1% |
Note: *These complications were assessed with severity ratings. Only the rates
for moderate, severe, or very severe (excludes mild and very mild ratings) are shown in
this table.
** As reported in original PMA submission.
The following table provides a breakdown of the types of surgical procedures that were performed through 4 and 5 years after implantation. Through 5 years, there were 224 patients who had one or more additional operations after the initial implantation (reoperations), for a total of 293 reoperations. These reoperations involved one or more surgical procedures, for a total of 463 surgical procedures. Examples of multiple procedures during a single reoperation include implant replacement for both breasts or a capsule procedure and mastopexy on the same breast. The most common type of additional surgical procedure through 5 years was implant removal with replacement (34% of the 463 procedures performed).
Type of Surgical Procedures |
N = 901 Patients |
|
Through 4 Years* |
Through 5 Years |
|
Implant Removal With Replacement** | 30% | 34% |
Capsule Procedure | 19% | 19% |
Adjust Saline Fill Volume | 11% | 11% |
Scar Revision/Wound Repair | 9% | 9% |
Removal of Excess Fluid | 7% | 6% |
Mastopexy | 7% | 6% |
Biopsy/Lump Removal | 4% | 5% |
Reposition Implant | 5% | 5% |
Implant Removal Without Replacement | 3% | 2% |
Removal of Skin Lesion/Cyst | 2% | 2% |
Exploration of Breast Area or Implants | 2% | 1% |
Skin Related Procedure | 1% | 1% |
Nipple Related Procedure (Unplanned) | <1% | <1% |
Total | 100% | 100% |
* As reported in orginal PMA submission with additional data clarification.
** Some removals were replaced with a McGhan Medical implant, while others
were replaced with a non-McGhan Medical implant.
The following table details the primary reasons for implant removal in the A95 Study over the 5 years. Through 5 years, there were 166 devices removed in 98 patients. Of these 166 devices, 156 were replaced and 10 were not. The most common reason for implant removal through 5 years was patient request for a size or style change (42% of the implants removed).
Primary Reason for Implant Removal |
Through 4 Years* |
Through 5 Years |
Patient Request for Size/Shape Change | 43% | 43% |
Leakage/Deflation | 33% | 33%** |
Capsular Contracture | 10% | 10% |
Wrinkling | 5% | 4% |
Implant Palpability/Visibility | 0% | 4% |
Asymmetry | 3% | 2% |
Breast Pain | 2% | 2% |
Implant Malposition | 2% | 1% |
Infection | 1% | 1% |
Implant Extrusion | 1% | 1% |
Damage to Implant During Surgery | 1% | 1% |
Total | 100% | 100% |
* As reported in original PMA submission with additional data clarification.
** Includes 1 implant removal where the reason for removal is unknown.
There were 78 patients in the A95 Study who had 126 implants removed and replaced with McGhan® implants. The table below reflects the number of replaced implants (not patients) out of 100 implants associated with the listed complications within 2 and 3 years following replacement. For example there was capsular contracture in 8% or 8 out of 100 implants at some time within 3 years after replacement. The complications reported following implant replacement were restricted to the same ones collected in the Large Simple Trial, LST (refer to page 11).
Complication
Following Replacement of Augmentation Implant(s) |
2-Year
Complication Rate* |
3-Year
Complication Rate* |
Removal/Replacement | 5% | 18% |
Leakage/Deflation | 9% | 9% |
Capsule Contracture III/IV | 7% | 8% |
Infection | 1% | 3% |
* As reported in original PMA submission.
Breast disease and connective tissue disease (CTD) were reported in some patients through 5 years after implantation in the A95 Study. Although there were 901 patients enrolled in the A95 Study, not every patient returned for each follow-up visit. Therefore, the percentage of patients with these events cannot be determined. Only the number of events can be reported. Without a comparison group of women with similar characteristics (such as age, race, etc.) and without breast implants, no conclusions can be made about the relationship between breast implants and these breast disease and CTD events.
There were 81 reports of breast disease through 5 years. Of these 81 reports, 7 are new since year 4. Additionally, 2 unknown outcome reports at year 4 were found to be false reports and were removed from the year 5 numbers, and the remaining 7 unknown outcome patients at year 4 were recategorized into benign. The reports of breast disease through 4 and 5 years are summarized in the following table.
Breast Disease Observation |
No. of Patient Reports |
No. of Patient Reports |
Benign | 66 | 80 |
Malignant | 1 | 1 |
Unknown Outcome | 7 | 0 |
* As reported in original PMA submission with additional data clarification: Benign includes 22 additional reports and unknown outcome (includes 2 fewer reports).
The Table below shows the number of patients reported to have CTD through 4 and 5 years after implantation. Confirmed reports were based on a diagnosis by a doctor. Unconfirmed reports were based on self-reports by the patients. There were 20 reports of CTD through 5 years. Of the 20 reports, 11 are new since year 4 (2 confirmed, 9 unconfirmed). Additionally, 3 unconfirmed reports at year 4 were found to be false reports and were removed from the year 5 numbers.
Connective Tissue Disease | Through 4 Years* | Through 5 Years | ||
No. of Confirmed Reports |
No. of Unconfirmed Reports |
No. of Confirmed Reports | No. of Unconfirmed Reports | |
Graves' Disease | 2 | 0 | 3 | 0 |
Hyperthyroiditis | 1 | 2** | 2 | 1 |
Lupos Erythematosus and/or Rheumatoid Arthritis | 0 | 3 | 0 | 1 |
Thyroiditis | 0 | 2 | 0 | 4 |
Chronic Fatigue Syndrome or Fibromyalgia | 2 | 0 | 2 | 4 |
Inflammatory Bowel Disease | 0 | 0 | 0 | 1 |
Raynaud's Phenomenon, Graves' Disease, Hyperthyroiditis, and Rheumatoid Arthritis |
0 | 0 | 0 | 1** |
Seronegative Spondylarthritis | 0 | 0 | 0 | 1 |
Total | 5 | 7 | 7 | 13 |
* As reported in original PMA submission.
** Patient was recategorized at 5 year timepoint.
The benefits of saline-filled breast implants in the A95 Study were assessed by a variety of outcomes, including bra cup size change, patient satisfaction, body image, body esteem, and self concept. These outcomes were assessed for patients with both original and replacement saline devices before implantation and at 3 years after surgery, except for bra size and satisfaction. Bra size was measured within the first year and a half after surgery. Satisfaction was measured at every follow-up visit through 5 years.
859 of the original 901 patients (95%) at 18 months were included in an analysis of cup size (5% did not provide data because pre/post measurements were not obtained or replacement/removal occurred prior to obtaining a post measurement). Of these 859 patients, the following shows the percentage of patients experiencing various changes in cup size:
683 of the original 901 patients (76%) were included in an analysis of satisfaction at 5 years (24% were not included because satisfaction data was not obtained or implant replacement/removal occurred prior to 5 years). Of these 683 patients, 95% indicated being satisfied with their breast implants at 5 years.
Before implantation, augmentation patients scored higher (better) than the general U.S. female population on the SF-36 and MOS-20 scales, which measure general health-related quality of life. After 3 years, augmentation patients showed a worsening in their SF-36 and MOS-20 scores. The following two scales showed no change over the 3 years: The Tennessee Self Concept Scale (which measures overall self concept) and The Body Esteem Scale (which measures overall self esteem related specifically to ones body). The Rosenberg Self Esteem Scale (which measures overall self esteem) showed a slight improvement over the 3 years. The Semantic Differential Scale (which measures attitudes about your breasts compared to attitudes about yourself) showed that patients experienced an increased positive attitude towards their breasts compared to themselves over the 3 years.
WHAT ARE THE ALTERNATIVES TO BREAST AUGMENTATION?
You are advised to wait a week after reviewing and considering this information before deciding whether to have augmentation surgery.
The following list of questions may help you to remind you of topics to discuss with your surgeon. You may have additional questions as well.
CHOOSING A SURGEON
When choosing an experienced surgeon who is experienced with breast implantation, you
should know the answers to the following questions:
Familiarize yourself with the following options in breast implant surgery and be prepared to discuss with your surgeon the following issues:
IMPLANT SHAPE AND SIZE
Depending on the desired shape you wish to achieve, you and your surgeon may choose a
round or contoured implant shape. Generally, the larger you want your cup size, the larger
the breast implant the surgeon will consider (measured in cubic centimeters, or
ccs). You should be aware that contoured implants that are placed submuscularly
(under your chest muscle) may assume a round shape after implantation.
Your surgeon will also evaluate your existing tissue to determine if you have enough to cover the breast implant. If you desire a breast implant size too large for your tissue, the surgeon may warn you that breast implant edges may be apparent or visible post-operatively. You may even risk surgical complications. Also, excessively large breast implants may speed up the effects of gravity and result in earlier droop or sag.
SURFACE TEXTURING
Textured surface implants were designed to reduce the chance of capsular contracture. Some
information in the literature with small numbers of patients suggests that surface
texturing reduces the chance of severe capsular contracture, but clinical information from
studies of a large number of women with INAMED implants shows no difference in the
likelihood of developing capsular contracture with textured implants compared to smooth
surfaced implants.
PALPABILITY
The following may cause implants to be more palpable (more easily felt): textured
implants, larger implants, subglandular placement, and the amount of skin/tissue available
to cover the implant.
IMPLANT PLACEMENT
The breast implant can be placed either partially under the pectoralis major
muscle (submuscular) or on top of the muscle and under the breast glands (subglandular)
depending on the thickness of your breast tissue and its ability to adequately
cover the breast implant. You should discuss with your surgeon the pros and
cons of the implant placement selected for you.
Breast before augmentation | Breast after subglandular augmentation |
Breast after submuscular augmentation |
The submuscular placement may make surgery last longer, may make recovery longer, may be more painful, and may make it more difficult to have some reoperation procedures than the subglandular placement. The possible benefits of this placement are that it may result in less palpable implants, less capsular contracture, and easier imaging of the breast with mammography.
The subglandular placement may make surgery and recovery shorter, may be less painful, and may be easier to access for reoperation than the submuscular placement. However, this placement may result in more palpable implants, more capsular contracture, and more difficult imaging of the breast with mammography.
INCISION
SITES
To permit the smallest possible incision, the implant is typically inserted
empty, and then filled with saline. You should discuss with your surgeon, the
pros and cons for the incision site specifically recommended for you.
There are three common incision sites: under the arm (axillary), around the nipple (periareolar), or within the breast fold (inframammary). If the incision is made under the arm, the surgeon may use a probe fitted with a miniature camera, along with minimally invasive (very small) instruments, to create a "pocket" for the breast implant.
Periareolar
This incision is most concealed, but is associated with a higher likelihood of inability
to successfully breast feed, as compared to the other incision sites.
Inframammary
This incision is less concealed than periareolar and associated with less
difficulty than the periareolar incision site when breast feeding.
Axillary
This incision is less concealed than periareolar and associated with less difficulty than
the periareolar incision site when breast feeding.
Umbilical/endoscopic
This incision site has not been studied and is not recommended.
SURGICAL SETTING AND ANESTHESIA
Augmentation surgery is usually performed on an outpatient basis, either in a hospital
operating room, surgery center, or surgical suite in the surgeons office. General
anesthesia is commonly used, and local anesthesia is also an option. The surgery usually
lasts one to two hours. Your surgeon will make an incision and create a pocket for the
breast implant. Then, the breast implant will be placed in the pocket, filled, and
positioned. Finally, the incision will be closed, usually with stitches, and possibly
taped.
POST-OPERATIVE CARE
You will probably feel somewhat tired and sore for several days following the operation,
and your breasts may remain swollen and sensitive to physical contact for a month or
longer. You may also experience a feeling of tightness in the breast area as your skin
adjusts to your new breast size.
Post-operative care may involve the use of a post-operative bra, compression bandage, or jog bra for extra support and positioning while you heal. At your surgeons recommendation, you will most likely be able to return to work within a few days, although you should avoid any strenuous activities that could raise your pulse and blood pressure for at least a couple of weeks. Your surgeon may also recommend breast massage exercises.
Note: If you experience fever, or noticeable swelling and/or redness in your implanted breast(s), you should contact your surgeon immediately.
INSURANCE
Insurance does not cover breast augmentation and may not cover reoperation (additional
surgery) and additional surgeons visits following augmentation.
The 3-year and 5-year complication rates are shown from the most common 5-year rate to the least common 5-year rate in the table below. The rates reflect the number of reconstruction patients out of 100 who experienced the listed complication at least once within the first 3 and 5 years after implantation. Some complications occurred more than once for some patients. The two most common complications experienced within the first 5 years of implantation were reoperation (45% or 45 patients out of 100) and asymmetry (39% or 39 patients out of 100).
Complications |
N = 237 Patients |
|
3-Year** Complication Rate |
5-Year Complication Rate |
|
Additional Operation (Reoperation) | 39% | 45% |
Assymmetry* | 33% | 39% |
Capsular Contracture | 25% | 36% |
Implant Replacement/Removal for Any Reason | 23% | 28% |
Implant Palpability/Visibility* | 20% | 27% |
Wrinkling* | 23% | 25% |
Breast Pain* | 15% | 18% |
Loss of Nipple Sensation* | 12% | 18% |
Implant Malposition* | 12% | 17% |
Leakage/Deflation | 6% | 8% |
Irritation/Inflammation* | 7% | 7% |
Intense Skin Sensation* | 6% | 6% |
Scarring Complications | 6% | 6% |
Infection | 5% | 6% |
Capsule Calcification* | 5% | 5% |
Seroma | 4% | 4% |
Skin/Tissue Necrosis | 4% | 4% |
Delayed Wound Healing* | 3% | 3% |
Implant Extrusion | 3% | 3% |
Skin Rash | 3% | 3% |
Hematoma | 1% | 1% |
Note:
* These complications were assessed with severity ratings. Only the rates for
moderate, severe, or very severe (excludes mild and very mild ratings) are shown
in this table.
** As reported in original PMA submission.
The following table provides a breakdown of the types of surgical procedures that were performed through 4 and 5 years after implantation. Through 5 years, there were 100 patients who had one or more additional operations after the initial implantation (reoperations), for a total of 126 reoperations. These 126 reoperations involved one or more surgical procedures, for a total of 159 surgical procedures. Examples of multiple procedures during a single reoperation include implant replacement for both breasts or a capsule procedure and mastopexy on the same breast. This table does not include the 282 planned nipple reconstruction and nipple/areolar tattoo procedures that occurred through 4 years. Only the 151 unplanned procedures through 4 years and the 159 unplanned procedures through 5 years are included in the table. The most common type of unplanned additional surgical procedure through 5 years was implant removal with replacement (31% of the 159 procedures performed).
Type of Surgical Procedures |
N = 237 Patients |
|
Through 4 Years* |
Through 5 Years |
|
Implant Removal With Replacement** | 30% | 31% |
Scar Revision/Wound Repair | 19% | 19% |
Implant Removal Without Replacement | 11% | 13% |
Capsule Procedure | 12% | 8% |
Adjust Saline Fill Volume | 6% | 6% |
Removal of Excess Fluid | 5% | 4% |
Reposition Implant | 4% | 4% |
Biopsy/Lump Removal | 5% | 4% |
Skin Related Procedure | 4% | 4% |
Nipple Related Procedures (unplanned) | 2% | 2% |
Other Procedures*** | 2% | 2% |
Removal of Skin Lesion/Cyst | 1% | 2% |
Total | 100% | 100% |
* As reported in original PMA submission with additional data clarification.
** Some removals were replaced with a McGhan Medical implant, while others
were replaced with a non-McGhan Medical implant.
*** Through 4 years, other procedures were liposuction, placement of
a stacked implant.
Through 5 years, other procedures were liposuction, placement of a stacked implant.
The following table details the primary reasons for implant removal in the R95 Study over the 5 years. Through 5 years, there were 70 devices removed in 62 patients. Of these 70 devices, 49 were replaced and 21 were not. The most common reason for implant removal through 5 years was capsular contracture (31% of the 70 implants removed).
Primary Reason for Implant Removal |
Through 4 Years* |
Through 5 Years |
Capsular Contracture III/IV | 26% | 31% |
Patient Request for Size/Shape Change | 23% | 21% |
Leakage/Deflation** | 16% | 17% |
Infection | 10% | 10% |
Implant Extrusion | 6% | 6% |
Implant Malposition | 6% | 4% |
Other*** | 3% | 4% |
Wrinkling | 3% | 3% |
Asymmetry | 3% | 1% |
Recurrent Breast Cancer | 3% | 1% |
Total | 100% | 100% |
* As reported in original PMA submission with additional data clarificaiton.
** Includes removals where the reason for removal was unknown.
*** Through 4 years, other reasons were abnormality of CT scan at mastectomy
site, poor tissue expansion due to radiation. Through 5 years, other reasons
were abnormality of CT scan at mastectomy site, poor tissue expansion due to
radiation, second stage breast reconstruction.
There were 37 patients in the R95 Study who had 40 implants removed and replaced with McGhan® implants. The table below reflects the number of replaced implants (not patients) out of 100 implants associated with the listed complications within 2 and 3 years following replacement. For example there was capsular contracture in 34% or 34 out of 100 implants at some time within 3 years after replacement. The complications reported following implant replacement were restricted to the same ones collected in the Large Simple Trial, LST (refer to page 11).
Complication
Following Replacement of Reconstruction Implant(s) |
2-Year Complication Rate* |
3-Year Complication Rate* |
Capsule Contracture III/IV | 33% | 34% |
Removal/Replacement | 26% | 27% |
Leakage/Deflation | 5% | 10% |
Infection | 7% | 3% |
* As reported in original PMA submission with correction to capsular contracture rate.
Breast disease and connective tissue disease (CTD) were reported in some patients through 5 years after implantation in the R95 Study. Although there were 237 patients enrolled in the R95 Study, not every patient returned for each follow-up visit. Therefore, the percentage of patients with these events cannot be determined. Only the number of events can be reported. Without a comparison group of women with similar characteristics (such as age, race, etc.) and without breast implants, no conclusions can be made about the relationship between breast implants and these breast disease and CTD events.
There were 99 reports of breast disease through 5 years. Of these 99 reports, 7 are new since year 4. Additionally, 1 unknown outcome report was recategorzed at 5 years. The reports of breast disease through 4 and 5 years are summarized in the following table.
Breast Disease Observation |
No. of Patient Reports |
No. of Patient Reports |
Benign | 72 | 75 |
Malignant | 19 | 24 |
Unknown Outcome | 1 | 0 |
* As reported in original PMA submission with additional data clarification:
Benign includes 61 additional reports.
The table below shows the number of patients reported to have CTD through 4 and 5 years after implantation. Confirmed reports were based on a diagnosis by a doctor. Unconfirmed reports were based on self-reports by the patients. There were 5 reports of CTD through 5 years. Of the 5 reports, 4 are new since year 4 (all unconfirmed). Additionally, 4 unconfirmed reports at year 4 were found to be false reports and were removed from the year 5 numbers.
Connective Tissue Disease | Through 4 Years* | Through 5 Years | ||
No. of Confirmed Reports |
No. of Unconfirmed Reports |
No. of Confirmed Reports | No. of Unconfirmed Reports | |
Graves' Disease | 1 | 0 | 1 | 0 |
Lupos Erythematosus and/or Rheumatoid Arthritis | 0 | 1 | 0 | 3 |
Inflammatory Bowel Disease | 0 | 1 | 0 | 0 |
Thyroiditis | 0 | 2 | 0 | 1 |
Total | 1 | 4 | 1 | 4 |
* As reported in original PMA submission.
The benefits of saline-filled breast implants in the R95 Study were assessed by a variety of outcomes, including patient satisfaction, body image, body esteem, and self concept. These outcomes were assessed for patients with both original and replacement saline devices before implantation and at 3 years after surgery, except for satisfaction which was measured at every follow-up visit through 5 years.
137 of the original 237 patients (58%) were included in an analysis of satisfaction at 5 years (42% were not included because satisfaction data was not obtained or implant replacement/removal occurred prior to 5 years). Of these 137 patients, 89% indicated being satisfied with their breast implants at 5 years.
Before implantation, reconstruction patients scored higher (better) than the general U.S. female population before implantation on some SF-36 scales, which measure general health-related quality of life. After 3 years, reconstruction patients showed an improvement in some of their SF-36 and MOS-20 scores. The following three scales showed no change over the 3 years: The Tennessee Self Concept Scale (which measures overall self concept), The Rosenberg Self Esteem Scale (which measures overall self esteem), and The Body Esteem Scale (which measures overall self esteem related specifically to ones body). The Semantic Differential Scale (which measures attitudes about your breasts compared to attitudes about yourself) showed that patients experienced an increased positive attitude towards their breasts compared to themselves over the 3 years.
SHOULD YOU HAVE BREAST RECONSTRUCTION?
Whether you decide to have breast reconstruction depends on your own individual case, medical condition, general health, lifestyle, emotional state, and breast size and shape. You may consider consulting your family, friends, breast implant support groups, and breast cancer support groups to help you in making this decision.
If you are considering breast reconstruction and do not have a plastic surgeon, ask your general surgeon for a referral for the names of experienced, board certified plastic surgeons in your area. Your general surgeon, plastic surgeon, and oncologist should work together to plan your mastectomy and reconstruction procedure to give you the best possible result.
WHAT ARE THE ALTERNATIVES TO BREAST RECONSTRUCTION?
You may choose not to undergo breast reconstruction. In this case, you may or may not decide to wear an external breast form (prosthesis) inside your bra. Breast forms are available in a variety of shapes, sizes, and materials such as foam, cotton, and silicone. Custom prostheses are also available to match the size and shape of your breast.
WHAT ARE THE CHOICES IN RECONSTRUCTIVE PROCEDURES?
The type of breast reconstruction procedure available to you depends on your medical situation, breast shape and size, general health, lifestyle, and goals. Women with small or medium sized breasts are the best candidates for breast reconstruction.
Breast reconstruction can be accomplished by the use of a prosthesis (a breast implant, either silicone gel or saline-filled), your own tissues (a tissue flap), or a combination of the two. A tissue flap is a section of skin, fat and/or muscle which is moved from your stomach, back or other area of your body, to the chest area, and shaped into a new breast.
Whether or not you have reconstruction with or without breast implants, you will probably undergo additional surgeries to improve symmetry and appearance. For example, because the nipple and areola are usually removed with the breast tissue in mastectomy, the nipple is usually reconstructed by using a skin graft from another area of the body or the opposite breast, in addition to tattooing the area. Nipple reconstruction is usually done as a separate outpatient procedure after the initial reconstruction surgery is complete.
BREAST RECONSTRUCTION WITH BREAST IMPLANTS
Your surgeon will decide whether your health and medical condition makes you an appropriate candidate for breast implant reconstruction. Women with larger breasts may require reconstruction with a combination of a tissue flap and an implant. Your surgeon may recommend breast implantation of the opposite, uninvolved breast in order to make them more alike (maximize symmetry) or he/she may suggest breast reduction (reduction mammoplasty) or a breast lift (mastopexy) to improve symmetry. Mastopexy involves removing a strip of skin from under the breast or around the nipple and using it to lift and tighten the skin over the breast. Reduction mammoplasty involves removal of breast tissue and skin. If it is important to you not to alter the unaffected breast, you should discuss this with your plastic surgeon, as it may affect the breast reconstruction methods considered for your case.
RECONSTRUCTION INCISION SITES
Most implants in breast reconstruction use the mastectomy scar either immediately (during the mastectomy procedure) or after tissue expansion.
SURGICAL SETTING AND ANESTHESIA
Reconstruction surgery is usually performed on an inpatient basis in an operating room. General anesthesia is most often used.
THE TIMING OF YOUR BREAST IMPLANT RECONSTRUCTION
The following description applies to reconstruction following mastectomy, but similar considerations apply to reconstruction following breast trauma or for reconstruction for congenital defects. The breast reconstruction process may begin at the time of your mastectomy (immediate reconstruction) or weeks to years afterwards (delayed reconstruction). Immediate reconstruction may involve placement of a breast implant, but typically involves placement of a tissue expander, which will eventually be replaced with a breast implant. It is important to know that any type of surgical breast reconstruction may take several steps to complete.
Two potential advantages to immediate reconstruction are that your breast reconstruction starts at the time of your mastectomy and that there may be cost savings in combining the mastectomy procedure with the first stage of the reconstruction. However, there may be a higher risk of complications such as deflation with immediate reconstruction, and your initial operative time and recuperative time may be longer.
A potential advantage to delayed reconstruction is that you can delay your reconstruction decision and surgery until other treatments, such as radiation therapy and chemotherapy, are completed. Delayed reconstruction may be advisable if your surgeon anticipates healing problems with your mastectomy, or if you just need more time to consider your options.
There are medical, financial and emotional considerations to choosing immediate versus delayed reconstruction. You should discuss with your surgeon, plastic surgeon, and oncologist, the pros and cons with the options available in your individual case.
SURGICAL CONSIDERATIONS TO DISCUSS WITH YOUR SURGEON
Discuss the advantages and disadvantages of the following options with your surgeon and your oncologist:
Immediate Reconstruction:
One-stage immediate reconstruction with a breast implant (implant only).
Two-stage immediate reconstruction with a tissue expander followed by delayed reconstruction several months later with a breast implant.
Delayed Reconstruction:
Two-stage delayed reconstruction with a tissue expander followed several months later by
replacement with a breast implant.
One-Stage Immediate Breast Implant Reconstruction
Immediate one-stage breast reconstruction may be done at the time of your mastectomy.
After the general surgeon removes your breast tissue, the plastic surgeon will then
implant a breast implant that completes the one-stage reconstruction. In reconstruction
following mastectomy, a breast implant is most often placed submuscularly.
Two-Stage (Immediate or Delayed) Breast Implant Reconstruction
Breast reconstruction usually occurs as a two-stage procedure, starting with
the placement of a breast tissue expander, which is replaced several months
later with a breast implant. The tissue expander placement may be done immediately,
at the time of your mastectomy, or be delayed until months or years later.
Side View, Breast Tissue Removed |
Side View, Expander Inserted and Filled |
STAGE 1: TISSUE EXPANSION
During a mastectomy, the general surgeon removes skin as well as breast tissue, leaving
the chest tissues flat and tight. To create a breast shaped space for the breast implant,
a tissue expander is placed under the remaining chest tissues.
The tissue expander is a balloon-like device made from elastic silicone rubber. It is inserted unfilled, and over time, sterile saline fluid is added by inserting a small needle through the skin to the filling port of the device. As the tissue expander fills, the tissues over the expander begin to stretch, similar to the gradual expansion of a womans abdomen during pregnancy. The tissue expander creates a new breast shaped pocket for a breast implant.
Tissue expander placement usually occurs under general anesthesia in an operating room. Operative time is generally one to two hours. The procedure may require a brief hospital stay, or be done on an outpatient basis. Typically, you can resume normal daily activity after two to three weeks.
Because the chest skin is usually numb from the mastectomy surgery, it is possible that you may not experience pain from the placement of the tissue expander. However, you may experience feelings of pressure, tightness or discomfort after each filling of the expander, which subsides as the tissue expands but may last for a week or more. Tissue expansion typically lasts four to six months.
STAGE 2: PLACING THE BREAST IMPLANT
After the tissue expander is removed, the unfilled breast implant is placed
in the pocket, and then filled with sterile saline fluid. In reconstruction,
following mastectomy, a breast implant is most often placed submuscularly. The
surgery to replace the tissue expander with a breast implant (implant exchange)
is usually done under general anesthesia in an operating room. It may require
a brief hospital stay or be done on an outpatient basis.
Post Mastectomy | Stage 1: Tissue Expander | Stage 2: Breast Implant and Nipple/Areola Reconstruction |
The breast can be reconstructed by surgically moving a section of skin, fat and muscle from one area of your body to another. The section of tissue may be taken from such areas as your abdomen, upper back, upper hip, or buttocks.
The tissue flap may be left attached to the blood supply and moved to the breast area through a tunnel under the skin (a pedicled flap), or it may be removed completely and reattached to the breast area by microsurgical techniques (a free flap). Operating time is generally longer with free flaps, because of the microsurgical requirements.
Flap surgery requires a hospital stay of several days and generally a longer recovery time than implant reconstruction. Flap surgery also creates scars at the site where the flap was taken and on the reconstructed breast. However, flap surgery has the advantage of being able to replace tissue in the chest area. This may be useful when the chest tissues have been damaged and are not suitable for tissue expansion. Another advantage of flap procedures over implantation is that alteration of the unaffected breast is generally not needed to improve symmetry.
The most common types of tissue flaps are the TRAM (transverse rectus abdominus musculocutaneous flap) (which uses tissue from the abdomen) and the Latissimus dorsi flap (which uses tissue from the upper back).
It is important for you to be aware that flap surgery, particularly the TRAM flap, is a major operation, and more extensive than your mastectomy operation. It requires good general health and strong emotional motivation. If you are very overweight, smoke cigarettes, have had previous surgery at the flap site, or have any circulatory problems, you may not be a good candidate for a tissue flap procedure. Also, if you are very thin, you may not have enough tissue in your abdomen or back to create a breast mound with this method.
THE TRAM FLAP (PEDICLE OR FREE)
Post Mastectomy | TRAM Flap | Final Result with Nipple/AreolaReconstruction |
During a TRAM flap procedure, the surgeon removes a section of tissue from your abdomen and moves it to your chest to reconstruct the breast. The TRAM flap is sometimes referred to as a "tummy tuck" reconstruction, because it may leave the stomach area flatter.
A pedicle TRAM flap procedure typically takes three to six hours of surgery under general anesthesia; a free TRAM flap procedure generally takes longer. The TRAM procedure may require a blood transfusion. Typically, the hospital stay is two to five days. You can resume normal daily activity after six to eight weeks. Some women, however, report that it takes up to one year to resume a normal lifestyle. You may have temporary or permanent muscle weakness in the abdominal area. If you are considering pregnancy after your reconstruction, you should discuss this with your surgeon. You will have a large scar on your abdomen and may also have additional scars on your reconstructed breast.
During a Latissimus Dorsi flap procedure, the surgeon moves a section of tissue from your back to your chest to reconstruct the breast. Because the Latissimus Dorsi flap is usually thinner and smaller than the TRAM flap, this procedure may be more appropriate for reconstructing a smaller breast.
Post Mastectomy | View Showing Back Scar | Latisimus Dorsi Flap and Nipple/Areola Reconstruction |
The Latissimus Dorsi flap procedure typically takes two to four hours of surgery under general anesthesia. Typically, the hospital stay is two to three days. You can resume daily activity after two to three weeks. You may have some temporary or permanent muscle weakness and difficulty with movement in your back and shoulder. You will have a scar on your back, which can usually be hidden in the bra line. You may also have additional scars on your reconstructed breast.
Depending on the type of surgery you have (i.e., immediate or delayed), the post-operative recovery period will vary.
Note: If you experience fever, or noticeable swelling and/or redness in your implanted breast(s), you should contact your surgeon immediately.
The following list of questions may help to remind you of topics to discuss with your surgeon. You may have additional questions as well.
CHOOSING A SURGEON
When choosing an experienced surgeon who is experienced with breast implantation, you
should know the answers to the following questions:
Familiarize yourself with the following options in breast implant surgery and be prepared to discuss with your surgeon the following issues:
IMPLANT SHAPE AND SIZE
Depending on the desired shape you wish to achieve, you and your surgeon may choose a
round or contoured implant shape. Generally, the larger you want your cup size, the larger
the breast implant the surgeon will consider (measured in cubic centimeters, or
ccs). You should be aware that contoured implants that are placed submuscularly may
assume a round shape after implantation.
Your surgeon will also evaluate your existing tissue to determine if you have enough to cover the breast implant. If you desire a breast implant size too large for your tissue, the surgeon may warn you that breast implant edges may be apparent or visible post-operatively. You may even risk surgical complications. Also, excessively large breast implants may speed up the effects of gravity and result in earlier droop or sag.
SURFACE TEXTURING
Textured surface implants were designed to reduce the chance of capsular contracture. Some
information in the literature with small numbers of patients suggests that surface
texturing reduces the chance of severe capsular contracture, but clinical information from
studies of a large number of women with McGhan® implants shows no difference in the
likelihood of developing capsular contracture with textured implants compared to smooth
surfaced implants.
PALPABILITY
The following may cause implants to be more palpable (more easily felt): textured
implants, larger implants, subglandular placement, and the amount of skin/tissue available
to cover the implant.
INSURANCE
Most insurance covers the first breast reconstruction operation. Insurance coverage for
reoperation procedures or additional surgeons visits following reconstruction may
not be covered, depending on the policy.
If you believe that you have experienced a serious problem(s) related to your breast implants, you should have your health professional report the problem(s) to FDA. You are encouraged to report any adverse events through your health professional. Although reporting by physicians or other health professionals is preferred, women may also report any serious problem directly through the MedWatch voluntary reporting system. An adverse event is serious and should be reported when it results in an initial or prolonged hospitalization, disability, congenital anomaly, or medical or surgical intervention. This information reported to MedWatch is entered into databases to be used to follow safety trends (patterns) of a device and to determine whether further follow-up of any potential safety issues related to the device is needed.
To report, use MedWatch form 3500 which may be obtained through FDAs website at http://www.fda.gov/medwatch/index.html. You may also call 1-888-463-INFO-FDA (1-888-463-6332), from 10:00am 4:00pm Eastern Time, Monday through Friday to receive an additional FDA MedWatch Package. Keep a copy of the MedWatch form completed by your surgeon for your records.
GENERAL RESOSURCES ABOUT IMPLANTS:
Upon request, you will be provided with a copy of the Directions for Use (package insert). You can request a copy from your surgeon or from INAMED Aesthetics. For more detailed information on the preclinical and clinical studies conducted by INAMED, you are referred to the Summary of Safety and Effectiveness Data for this product at http://www.fda.gov/cdrh/pdf/p990074.html.
You will be given a device identification card with the style and serial number of your breast implant(s).
INAMED Aesthetics
1-800-624-4261
www.inamed.com
Institute of Medicine Report on the Safety of Silicone Implants
www.nap.edu/catalog/9618.html
Food and Drug Administration
1-888-INFO-FDA or 301-827-3990
www.fda.gov/cdrh/breastimplants/
BREAST RECONSTRUCTION RESOURCES
The following list of resources may help you to find more information and support for your breast reconstruction decision.
National Cancer Institute
1-800-4-CANCER
www.cancernet.nci.nih.gov
American Cancer Society (Reach to Recovery)
1-800-ACS-2345
www.cancer.org
Y-ME National Organization for Breast Cancer Information and Support
1-800-221-2141
www.y-me.org
Areola | The pigmented or darker colored area of skin surrounding the nipple of the breast. |
Asymmetry |
A lack of proportion of shape, size and position on opposite sides of the body. |
Autoimmune Disease |
A disease in which the body mounts an "attack" Disease response to its own tissues or cell types. Normally, the bodys immune mechanism is able to distinguish clearly between what is a normal substance and what is foreign. In autoimmune diseases, this system becomes defective and produces antibodies against normal parts of the body, causing tissue injury. Certain diseases such as rheumatoid arthritis and scleroderma are considered to be autoimmune diseases. |
Axillary |
Pertaining to the armpit area. |
Bilateral |
Pertaining to both the left and right breast. |
Biopsy |
Removal and examination of sample tissue for diagnosis. |
Breast Augmentation |
Enlargement of the breast by surgical implantation of a breast implant or patients own tissue. |
Breast Reconstruction |
Surgical restoration of natural breast contour and mass following mastectomy, trauma or injury. |
Capsular Contracture |
Tightening of the tissue surrounding a breast implant which results in a firmer breast. |
Capsulectomy |
Surgical removal of the entire capsule surrounding a breast implant. |
Capsulotomy |
Closed Capsulotomy: Compression on the outside of the breast to break the capsule and relieve contracture. Open Capsulotomy: Surgically cutting or removing part of the capsule through an incision. |
Carcinoma |
Invasive malignant tumor. |
Congenita Anomaly |
Abnormality existing at birth. |
Connective Tissue Disease (CTD) |
A disease or group of diseases affecting connective tissue. The cause of these diseases are unknown. The diseases are grouped together on the basis of clinical signs, symptoms, and laboratory abnormalities. |
Deflation/ Rupture |
Refers to loss of saline from a saline-filled breast implant due to a tear or cut in the implant shell or possibly a valve leak. |
Displacement |
Shifting in the original position. |
Epidemiological |
Pertaining to the cause, distribution and control of disease in populations. |
Extrusion |
A breast implant or tissue expander being pressed out of the body. |
Fibrous Tissue |
Tissue resembling fibers. |
Hematoma |
A swelling or mass of blood (usually clotted) confined to an organ, tissue, or space and caused by a break in a blood vessel. |
Immune Response |
The reaction of the body to substances that areforeign or are interpreted as being foreign. |
Inframammary |
Below the breast. |
Inframammar Fold |
The crease at the base of the breast and the chest wall. |
Inframammary Incision |
A surgical incision at the inframammary fold. |
In-Patient Surgery |
Surgery performed in a hospital requiring an overnight stay |
Latissimus Dorsi |
Two triangular muscles running from the spinal column to the shoulder. |
Mammaplasty |
Plastic surgery of the breast. |
Mammary |
Pertaining to the breast. |
Mammography |
Use of radiography (X-rays) of the breast to detect breast cancer. Recommended as a screening technique for early detection of breast cancer. |
Mastectomy |
Surgical removal of the breast. Subcutaneous Mastectomy: Removal of breast tissue, preserving the skin and nipple. Partial Mastectomy: Removal of primary tumor and a wide margin of tissue, may include the overlying skin and the muscle fibrous tissue (fascia) underlying the tumor. Total (Simple) Mastectomy: Removal of breast tissue and the nipple; sometimes accompanied by armpit (axillary) node dissection. Modified Radical Mastectomy: Removal of breast tissue, nipple, and fascia of chest (pectoralis) muscle with axillary node dissection. |
Mastopexy |
Plastic surgery to move sagging (ptotic) breasts into a more elevated position. |
Necrosis |
Death of tissue. May be caused by insufficient blood supply, trauma, radiation, chemical agents or infectious disease. |
Oncologist |
A specialist in the branch of medicine dealing with the study and treatment of tumors. |
Out-Patient Surgery |
Surgery performed in a hospital or surgery Surgery center not requiring an overnight stay. |
Palpate/ Palpability |
To feel with the hand. |
Pectoralis |
The major muscle of the chest. |
Plastic Surgery |
Surgery intended to improve, restore, repair, or reconstruct portions of the body following trauma, injury or illness. |
Prosthesis |
An artificial device used to replace or represent a body part. |
Ptosis |
Sagging of the breast usually due to normal aging, pregnancy or weight loss.. |
Rectus Abdominus |
Major abdominal (stomach) muscle. |
Saline |
A solution of sodium chloride (salt) and water. |
Seroma |
Localized collection of serum, the watery portion of blood, that resembles a tumor. |
Serratus |
Muscle located beneath the chests pectoralis major and minor muscles and the rib cage. |
Silicone Elastomer |
A type of silicone that has elastic properties similar to rubber. |
Subglandular Placement |
Placement of the breast implant behind the skin and mammary gland, but on top of the chest (pectoralis) muscle. Also called prepectoral or retromammary placement. |
Submuscular Placement |
Placement of the breast implant under the chest (pectoralis) muscle, or under the pectoralis and serratus muscles. Also called retropectoral or subpectoral placement. |
Surgical Incision |
Cut made in tissue for surgical purposes. |
Transaxillary Incision |
Incision across the long axis of the armpit (axilla). |
Umbilical |
Relating to the navel. |
Unilateral |
Affecting only left or right breast. |