FDA
Home Page | CDRH Home Page | Search
| CDRH
A-Z Index | Contact CDRH
|
PMA Final Decisions Rendered for January 2001 |
A PDF document that contains the "Approval letter and Summary of Safety and Effectiveness" is being added to this listing for each PMA. The PMA number will appear as a link if this document is available. Information about the PDF Reader is available.
APPLICATION NUMBER / DATE of APPROVAL | DEVICE TRADE NAME | COMPANY NAME CITY, STATE, & ZIP | DEVICE DESCRIPTION / INDICATIONS |
1/24/01 |
SUPARTZ Dispo |
Seikagaku Corp. Tokyo, Japan |
Approval for the SUPARTZ Dispo. The device is indicated for the treatment of pain in osteoarthritis (OA) of the knee in patients who have failed to respond adequately to conservative non-pharmacologic therapy and simple analgesics, e.g., acetaminophen. |
1/5/01 |
TMJ Fossa-Eminence/Condylar Prosthesis System |
TMJ Implants, Inc. Golden, CO 80401 |
Approval for the TMJ Fossa-Eminence/Condylar Prosthesis System. The device is indicated for reconstruction of the temporomandibular joint. Patients should be considered if they have one or more of the following conditions: 1) Inflammatory arthritis involving the temporomandibular joint not responsive to other modalities of treatment 2) Recurrent fibrous and/or bony ankylosis not responsive to other modalities of treatment 3) Failed tissue graft 4) Failed alloplastic joint reconstruction 5) Loss of vertical mandibular height and/or occlusal relationship due to bone resorption, trauma, developmental abnormality, or pathologic lesion. |
APPLICATION NUMBER / DATE of APPROVAL |
DEVICE TRADE NAME |
COMPANY NAME |
DEVICE DESCRIPTION / INDICATIONS |
P790017/S074 1/8/01 Real-Time |
D114S Over-The-Wire Balloon Dilatation Catheter |
Medtronic AVE, Inc. Santa Rosa, CA 95403 |
Approval for the addition of six (6) catalogue items to the approved D114S Over-The-Wire Balloon Dilatation Catheter product line. These six items consist of a 10 mm balloon length in each of the current balloon diameters of 1.5 mm 2.0 mm, 2.5 mm, 3.0 mm, 3.5 mm, and 4.0 mm. In addition, approval for a minor dimensional change to the 1.5 mm diameter extruded balloon tubing (0.0135"/0.0205" ID/OD from the current 0.014"/0.022" ID/OD). The device is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion. |
P810046/S203 1/11/01 Real-Time |
RX CHASSIS 1 Coronary Dilatation Catheter |
Guidant Corporation Temecula, CA 92589 |
Approval for the RX CHASSIS 1 Coronary Dilatation Catheter. The device, as modified, is indicated for a) balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion, and b) balloon dilatation of a coronary artery occlusion for the purpose of restoring coronary flow in patients with ST segment elevation myocardial infarction. |
P810046/S205 1/24/01 Real-Time |
CrossSail and OpenSail Coronary Dilatation Catheters |
Guidant Corporation Temecula, CA 92589 |
Approval for the revision of the CrossSail and OpenSail Coronary Dilatation Catheter’s compliance charts. These catheters are indicated for: 1) balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion and 2) balloon dilatation of a coronary artery occlusion for the purpose of restoring coronary flow in patients with ST-segment elevation myocardial infarction. |
P840024/S080 1/16/01 180-Day |
Nucleus 22 Cochlear Implant System - Adults |
Cochlear Corporation Englewood, CO 80112 |
Approval for modification to the firmware, components and labeling of the Portable Programming System (PPS) with no change to the currently approved indications for use. |
P850089/S045 1/8/01 Real-Time |
Medtronic® CapSure® Novus Steroid Eluting, Bipolar, Transvenous Pacing Lead Models 5054 and 5554. |
Medtronic, Inc. Shoreview, MN 55126 |
Approval to distribute the Medtronic® CapSure® Z Novus Models 5054 and 5554 pacing leads under a new trade name. The devices will be marketed under the trade name Vitatron Impulse II Models IHP 09B and IHP 09JB and are indicated for providing permanent pacing and sensing of the ventricle (Model IHP 09B) and atrium (Model IHP 09JB) when used with a compatible pulse generator. |
P860003/S037 1/5/01 180-Day |
UVAR® XTS Photopheresis System |
Therakos Exton, PA 19341 |
Approval for the following software changes: 1) removal of system error F522:0000, 2) removal of system error F581:0000, 3) change in the alarm threshold for the centrifuge bowl vacuum, and 4) addition of new priming state. |
P860019/S167 1/11/01 Real-Time |
Boston Scientific Scimed® Quantum Monorail Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter |
Boston Scientific Scimed, Inc. Maple Grove, MN 55311 |
Approval for the Boston Scientific Scimed Quantum Monorail Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter (9, 12, 15, 18, and 20 mm lengths with balloon diameters of 2.5 – 4.0 mm in ¼ mm increments). The quantum Monorail PTCA Catheter is indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion. The Quantum Monorail PTCA Catheter (2.5 – 4.0 mm balloon diameters) is also indicated for the post-delivery expansion of balloon expandable stents. |
P860026/S007 1/26/01 Real-Time |
Diaphragmatic /Phrenic Nerve Stimulator |
Avery Laboratories, Inc. Commack, NY 11725 |
Approval for changing the method of attaching the battery connector and wiring from the power switches to the printed circuit board, changing the point where the signal is taken to illuminate the battery LED indicator, adding the energy source routing to the front panel of the respiratory rate control circuitry, changing the point where the battery power is obtained for each stimulator output amplifier circuit, and decreasing the sensitivity of the oscillator circuit to transistor output capacitance. The device, as modified, will be marketed under the trade name Mark IV and is indicated for persons who require chronic ventilatory support because of upper motor neuron respiratory muscle paralysis (RMP) or because of central alveolar hypoventilation (CAH) and whose remaining phrenic nerve, lung, and diaphragm function is sufficient to accommodate electrical stimulation. |
P890027/S049 1/16/01 180-Day |
Nucleus 22 Cochlear Implant System - Children |
Cochlear Corporation Englewood, CO 80112 |
Approval for modification to the firmware, components and labeling of the Portable Programming System (PPS) with no change to the currently approved indications for use. |
P910073/S037 1/23/01 Real-Time |
Model 6942 Bi-directional Torque Wrench |
Guidant/CRM St. Paul, MN 55112 |
Approval for modifications to the Model 6942 Bi-directional Torque Wrench. |
P910077/S036 1/23/01 Real-Time |
Model 6942 Bi-directional Torque Wrench |
Guidant/CRM St. Paul, MN 55112 |
Apporoval for modifications to the Model 6942 Bi-directional Torque Wrench. |
P930021/S004 1/19/01 Real-Time |
EMDOGAIN® Gel |
Biora, Inc. Chicago, IL 60610 |
Approval to change from the EMDOGAIN® two vial system to the EMDOGAIN® Gel filled sterile syringes. The change also includes the addition of the following new manufacturing sites: Statens Serum Institut, Copenhagen, Denmark; Orifice Medical AB, Ystad, Sweden; Paper-Pak Sweden AB, Aneby, Sweden. The device, as modified, will be marketed under the trade name EMDOGAIN® Gel. The device is indicated for use as an adjunct to periodontal surgery as a topical application onto exposed root surfaces to treat intrabony defects not involving furcations due to moderate or severe periodontitis. The device is also indicated for use as an adjunct to a minimally invasive surgical technique in esthetic zones to optimize tissue height for intrabony defects only. |
P930035/S009 1/23/01 Real-Time |
Model 6942 Bi-directional Torque Wrench |
Guidant/CRM St. Paul, MN 55112 |
Approval for modifications to the Model 6942 Bi-directional Torque Wrench. |
P940022/S016 1/12/01 180-Day |
Clarion® Multi-Strategy Cochlear Implant Platinum BTE Sound Processor |
Advanced Bionics Corporation Sylmar, CA 91342 |
Approval for incorporation of the size-reduced speech processor to be worn behind the ear, known as the platinum BTE Sound Processor. The device, as modified, will be marketed under the trade name Clarion® Multi-Strategy Cochlear Implant Platinum BTE Sound Processor. The device is indicated to restore a level of auditory sensation to individuals with profound sensorineural deafness via electrical stimulation of the auditory nerve. Adults: 1) 18 years of age of older 2) Profound, bilateral sensorineural deafness (³ 90 dB) 3) Postlingually deafened 4) Lack of benefit from appropriately fitted hearing aids. |
P950015/S008 1/29/01 180-Day |
Heart Laser 2 (HL2) |
PLC Medical Systems, Inc. Franklin, MA 02038 |
Approval for a manufacturing change to up-date the laser console portion of the device system from the currently approved HL1 (Heart Laser) to the HL2 (Heart Laser 2). The device, as modified, will be marketed under the trade name Heart Laser 2 (HL2) , and is indicated for use in the treatment of patients with stable angina (Canadian Cardiovascular Society class 3 or 4) refractory to medical treatment and secondary to objectively demonstrated coronary artery atherosclerosis and with a region of the myocardium not amenable to direct coronary revascularization. |
P960025/S004 1/12/01 Real-Time |
Lumbar I/F Cage® System |
DePuy AcroMed Raynham, MA 02767 |
Approval for the use of the Lumbar I/F cage components of the Lumbar I/F Cage® With VSP® Spine System with additional pedicle screw spinal systems. Specifically, for use with the following six additional posterior pedicle screw spinal systems: 1) the Ti alloy version of the VSP Spine System; 2) Ti alloy and stainless steel versions of the ISOLA Spine System; 3) TiMX Spine System; and 4) the Ti alloy and stainless steel versions of the MOSS Miami Spine System. In addition, the device, as modified, will be marketed under the trade name Lumbar I/F Cage System. The indication for these implants has not been changed from that approved on February 2, 1999. |
P960040/S019 1/23/01 Real-Time |
Model 6942 Bi-directional Torque Wrench |
Guidant/CRM St. Paul, MN 55112 |
Approval for modifications to the Model 6942 Bi-directional Torque Wrench. |
P960042/S006 1/17/01 180-Day |
Spectranetics Laser Sheath Kits, 12 French, 14 French, and 16 French, Models 500-001, 500-012, and 500-013 |
Spectranetics Corporation Colorado Springs, CO 80907 |
Approval for an increase in the laser "on" time to 10 seconds and a decrease in the laser "off" time to 5 seconds. These devices are indicated for use as an adjunct to conventional lead extraction tools in patients suitable for transvenous removal of chronically implanted pacing or defibrillator leads constructed with silicone or polyurethane outer insulation. |
P960058/S015 1/12/01 180-Day |
Clarion® Multi-Strategy Cochlear Implant Platinum BTE sound Processor |
Advanced Bionics Corporation Sylmar, CA 91342 |
Approval for incorporation of the size-reduced speech processor to be worn behind the ear, known as the platinum BTE Sound Processor. The device, as modified, will be marketed under the trade name Clarion® Multi-Strategy Cochlear Implant Platinum BTE Sound Processor. The device is indicated to restore a level of auditory sensation to individuals with profound sensorineural deafness via electrical stimulation of the auditory nerve. Children: 1) Two years of age through 17 years, if X-rays demonstrate evidence of ossification, children as young as 18 months may be implanted. 2) Profound, bilateral sensorineural deafness (³ 90 dB) 3) Undergone or be willing to undergo a hearing aid trial with appropriately fitted hearing aids. 4) Lack of benefit from appropriately fitted hearing aids. In younger children, lack of benefit with hearing aids is defined as failure to attain basic auditory milestones such as a child’s inconsistent response to his/her name in quiet or to environmental sounds (Meaningful Auditory Integration Scale). In older children, lack of aided benefit is defined as scoring 0% on open-set work recognition (Phonetically Balanced Kindergarten test-Word List) administered with monitored live-voice (70 dB SPL). Both younger and older children should demonstrate only minimal ability on age appropriate open-set sentence measures and a plateau in auditory development. |
D970003/S013 1/22/01 Real-Time |
DISCOVERY II /INTELIS II Pacemakers and CONSULT Model 2891 Software Application |
Guidant/CRM St. Paul, MN 55112 |
Approval for modifications to the CONSULT Model 2891 Software Application for the DISCOVERY II /INTELIS II Pacemakers, resulting in version 2.4 of the CONSULT programmer software. |
D970003/S014 1/22/01 Real-Time |
Pulsar Max Models 1170 and 1270 |
Guidant/CRM St. Paul, MN 55112 |
Approval for modifications to the Pulsar Max Models 1170 and 1270, specifically the creation of version having headers which are "IS-1 only" compatible. |
P970034/S007 1/17/01 180-Day |
UV-Absorbing Posterior Chamber Intraocular Lenses |
Ophthalmic Innovations International, Inc. Claremont, CA 91711 |
Approval for a new manufacturing facility located at Ophthalmic Innovations International, Inc., Ontario, California. |
P970051/S012 1/16/01 180-Day |
Nucleus 24 Cochlear Implant System for Adults and Children |
Cochlear Corporation Englewood, CO 80112 |
Approval for modification to the firmware, components and labeling of the Portable Programming System (PPS) with no change to the currently approved indications for use. |
P970058/S009 1/22/01 180-Day |
ImageChecker M1000 |
R2 Technology, Inc. Los Altos, CA 94022 |
Approval for the addition of a new calcification marker. This marker (a circle) is intended to enable the user of the ImageChecker to differentiate obvious calcifications from more subtle calcifications that appear in analyzed mammograms. |
P980001/S023 1/25/01 180-Day |
NIROYAL Elite Premounted Stent System |
Boston Scientific Scimed, Inc. Maple Grove, MN 55311 |
Approval for additional stent lengths (i.e., 9 mm, 12 mm and 18 mm) for the NIROYAL Elite Premounted Stent System. The device is indicated for improving coronary luminal diameter in the following: 1) patients with symptomatic ischemic disease associated with stenotic lesions in native coronary arteries (length £ 25 mm) in reference vessel diameter of 3.0 to 4.0 mm; 2) treatment of abrupt or threatened closure in patients with failed interventional therapy in lesions with reference diameters in the range of 2.5 to 4.0 mm; and 3) patients with symptomatic heart disease due to lesions in saphenous vein bypass grafts with lesion length £ 30 mm and reference diameter in the range of 3.0 to 4.0 mm. |
P980022/S006 1/9/01 180-Day |
MiniMed Continuous Glucose Monitoring System |
MiniMed, Inc. Sylmar, CA 91342 |
Approval for a new manufacturing site for the manufacturer of the external components of the device. The new site is MiniMed, Inc., Northridge, California. |
P980049/S004 1/31/01 Real-Time |
Orchestra Programmer |
ELA Medical Plymouth, MN 55441 |
Approval for a "Caution:" label to be placed on the Orchestra Programmer. |
P990001/S005 1/8/01 Real-Time |
Vitatron Diva Platform |
Vitatron, Inc. Minneapolis, MN 55459 |
Approval for an extension of the shelf life date to 18 months from 12 months and related labeling changes. |
P990009/S001 1/22/01 Real-Time |
Proceed Hemostatic Sealant |
Fusion Medical Technologies, Inc. Fremont, CA 94555 |
Approval for the removal of the neurosurgical exclusion from the "Indications for Use" for Proceed Hemostatic Sealant. |
P990017/S009 1/19/01 180-Day |
ANCURE System |
Guidant Corporation Menlo Park, CA 94025 |
Approval for modifications to the ANCURE System Instructions for Use. |
P990030/S001 1/2/01 Real-Time |
CoStasis®/ DynaStat Surgical Hemostat |
Cohesion Technologies Palo Alto, CA 94303 |
Approval for a change in the method of terminal sterilization from E-Beam to gamma irradiation and requested an additional 3-ml delivery syringe device with the associated labeling change. |
P990052/S002 1/3/01 180-Day |
Vibrant® D Soundbridge System (Audio Processor D Model 404) |
Symphonix Devices, Inc. San Jose, CA 95131 |
Approval for the incorporation of an alternate source supplier’s signal processing circuit into the Audio Processor D (Model 404). The device is indicated for adults, 18 years of age or older, who have a moderate to severe sensorineural hearing loss and desire an alternative to an acoustic hearing aid. Prior to receiving the device, it is recommended that an individual have experience with appropriately fit hearing aids. |
P990066/S002 1/8/01 180-Day |
Senographe 2000D Full Field Digital Mammography System |
General Electric Medical Systems Milwaukee, WI 53201 |
Approval for the Kodak DryView 8610 Printer for use with the Senographe 2000D Full Field Digital Mammography System. |
30-DAY NOTICES (135 Day Supplement Was Not Required) | |||
APPLICATION NUMBER & LTR DATE | DEVICE TRADE NAME | COMPANY NAME CITY, STATE, & ZIP | COMMENTS |
P870038/S004 1/5/01 |
Starr Edwards Silastic Ball Valve Models 1260 and 6120 |
Edward Lifesciences LLC Irvine, CA 92614 |
Change in the packaging design from a two-piece molded polysulfone canister to a thermoformed polycarbonate tray with insert and Tyvek lid. |
P910023/S050 1/18/01 |
Cadence Family of Implantable Cardioverter Defibrillators |
St. Jude Medical, Inc. Sunnyvale, CA 94086 |
Use of an alternate sterilization chamber for the ICDs and tachycardia leads and is also requesting approval for use of parametric release as an alternate to the current use of biological indicators in the release of sterilized products. |
P930038/S028 1/16/01 |
Angio-Seal Vascular Closure Device |
St. Jude Medical, Inc. Minnetonka, MN 55345 |
Change from a manual crimping procedure to an automated pneumatic process to seal the crimp stop onto the suture. |
P950022/S012 1/18/01 |
Cadence Family of Implantable Cardioverter Defibrillators |
St. Jude Medical, Inc. Sunnyvale, CA 94086 |
Use of an alternate sterilization chamber for the ICDs and tachycardia leads and is also requesting approval for use of parametric release as an alternate to the current use of biological indicators in the release of sterilized products. |
P980053/S002 1/23/01 |
Durasphere Injectable Bulking Agent® |
Advanced UroScience Saint Paul, MN 55110 |
Addition of an alternate vendor, Carbon Coating Technologies, to supply the pyloric carbon coated bead component of the Durasphere Injectable Bulking Agent® |
P990025/S003 1/11/01 |
Navi-Star Diagnostic/Ablation Deflectable Tip Catheter |
Biosense Webster, Inc. Diamond Bar, CA 91765 |
Change to the polymeric mixture used to manufacture the catheter’s tip. The change is intended to standardize processes across the Navi-Star product line. |
P000018/S006 1/29/01 |
Novoste Beta-Cath System |
Novoste Corporation Norcross, GA 30093 |
Change in the insert molding process parameters for a component of the proprietary connector (the puck), which is molded to the 3-lumen tubing of the delivery catheter. |
Summary of PMA Originals & Supplements Approved
Originals: 2
Supplements: 31
Summary of PMA Originals Under Review
Total Under Review: 75
Total Active: 34
Total On Hold: 41
Number Greater Than 180 Days: 0
Summary of PMA Supplements Under Review
Total Under Review:235
Total Active: 152
Total On Hold: 83
Number Greater Than 180 Days: 5
Summary of All PMA Submissions Received
Originals: 1
Supplements: 55
Summary of PMA Supplement PMA Approval/Denial Decision Times
Number of Approvals: 2
Number of Denials: 0
Average Days Fr Receipt to Decision (Total Time): 94
FDA Time: 71.7 Days MFR Time: 22.3 Days
Revised 4/1/2002
CDRH Home Page | CDRH A-Z Index | Contact CDRH | Accessibility | Disclaimer
FDA Home Page | Search FDA Site | FDA A-Z Index | Contact FDA | HHS Home Page
Center for Devices and Radiological Health / CDRH