The current Good Manufacturing Practice (GMP) requirements set forth in the Quality
System (QS) regulation are promulgated under section 520 of the Food, Drug and Cosmetic
(FD&C) Act. They require that domestic or foreign manufacturers have a quality system
for the design, manufacture, packaging, labeling, storage, installation, and servicing of
finished medical devices intended for commercial distribution in the United States. The
regulation requires that various specifications and controls be established for devices;
that devices be designed under a quality system to meet these specifications; that devices
be manufactured under a quality system; that finished devices meet these specifications;
that devices be correctly installed, checked and serviced; that quality data be analyzed
to identify and correct quality problems; and that complaints be processed. Thus, the QS
regulation helps assure that medical devices are safe and effective for their intended
use. The Food and Drug Administration (FDA) monitors device problem data and inspects the
operations and records of device developers and manufacturers to determine compliance with
the GMP requirements in the QS regulation.
The QS Regulation is contained in Title 21 Part
820 of the Code of Federal Regulations. This regulation covers quality management
and organization, device design, buildings, equipment, purchase and handling
of components, production and process controls, packaging and labeling control,
device evaluation, distribution, installation, complaint handling, servicing,
and records. The preamble describes the public comments received during the
development of the QS regulation and describes the FDA Commissioner's resolution
of the comments. Thus, the preamble contains valuable insight into the meaning
and intent of the QS regulation.
The Good Manufacturing Practice (GMP) / Quality System Regulation page has a link to the Medical Device Quality Systems Manual: A Small Entity Compliance Guide, First Edition which details the requirements of the new QS regulation and provides detailed guidance in the following areas:
Manufacturers should use good judgment when developing their quality system and apply
those sections of the QS regulation that are applicable to their specific products and
operations, 21 CFR 820.5 of the QS regulation. Operating within this flexibility, it is
the responsibility of each manufacturer to establish requirements for each type or family
of devices that will result in devices that are safe and effective, and to establish
methods and procedures to design, produce, and distribute devices that meet the quality
system requirements. FDA has identified in the QS regulation the essential elements that a
quality system shall embody for design, production and distribution, without prescribing
specific ways to establish these elements. Because the QS regulation covers a broad
spectrum of devices and production processes, it allows some leeway in the details of
quality system elements. It is left to manufacturers to determine the necessity for, or
extent of, some quality elements and to develop and implement specific procedures tailored
to their particular processes and devices. For example, if it is impossible to mix up
labels at a manufacturer because there is only one label or one product, then there is no
necessity for the manufacturer to comply with all of the GMP requirements under device
labeling.
The medical device QS regulation requires an "umbrella" quality system
intended to cover the design, production, and distribution of all medical devices from
simple surgical hand tools to very complex computerized axial tomography (CAT) scanners.
It is not practical for a regulation to specify details of quality system elements for
such a wide range of products. Rather, the QS regulation specifies general objectives such
as use of trained employees, design reviews, design validation, calibrated equipment,
process controls, etc., rather than methods, because a specific method would not be
appropriate to all operations.
In most cases, it is left to the manufacturer to determine the best methods
to attain quality objectives. In some cases, however, the QS regulation does
specify the particular type of method to be used, such as written procedures
or written instructions. This does not mean, however, that manufacturers cannot
vary from the method specified if the intent of the GMP requirement can be met
by another method such as using an engineering drawing plus a model device as
manufacturing instructions. Written procedures are not restricted to paper copies.
Written procedures may be filed and distributed by automated data processing
equipment. This flexibility is allowed by section 21
CFR 820.180.
Typically, large manufacturers will have a quality system that exceeds the medical
device QS regulation. Small manufacturers will typically have a proportionally simpler
system. FDA recognizes that a small manufacturer may not need the same amount of
documentation that a large manufacturer does in order to achieve a stateofcontrol;
and, that some of the records maintained to fulfill the GMP requirements for written
procedures may not be as long and complex for a small manufacturer.
After a manufacturer establishes a quality system, it should be maintained. Each manufacturer should assure that with growth and process or product changes their quality system is still adequate. This assurance is obtained through change control, daytoday observance of operations, and by periodic audits of the quality system. The auditor should first identify the elements of the company's quality system. Next the audit should determine how well each element is functioning, and then determine its adequacy with respect to the intent of the device GMP requirements and meeting the company's quality claims.
The QS regulation applies to finished devices intended to be commercially distributed
for human use unless there is an approved exemption in effect. GMP exemptions
are codified in the Classification Regulations 21 CFR 862 to 892. The exemption
of most Class I devices from design controls is in section 21
CFR 820.30 (a).
Certain components such as blood tubing and major diagnostic xray components are
considered by FDA to be finished devices because they are accessories to finished devices.
The manufacturer of such accessories is subject to the QS regulation when the accessory
device is labeled and sold separately from the primary device for a healthrelated
purpose to a hospital, physician, or other user.
The designation of a device as a "custom" or "customized" device does not confer a GMP exemption.
Contract manufacturers and specification developers shall comply with the sections of
the QS regulation that apply to the functions they perform.
Contract test laboratories are considered an extension of a manufacturer's
quality system and presently are not routinely scheduled for GMP inspections.
The finished device manufacturer shall meet the requirement of the QS regulation,
particularly 21
CFR 820.50, Purchasing, when they obtain products or services. Internal
test laboratories, however, that are part of a corporate manufacturer that provides
services to individual corporation factories should meet GMP requirements. Internal
laboratories are inspected as part of the FDA GMP inspection of the member factories.
Situations are discussed in the remainder of this chapter where various manufacturers are exempt from the QS regulation or are not routinely inspected. However, these manufacturers are still subject to the FD&C Act. If these manufacturers or any manufacturer render devices unsafe or ineffective, the devices are adulterated and/or misbranded and the manufacturers are subject to the penalties of the FD&C Act.
FDA has determined that certain types of establishments are exempt from GMP
requirements; and FDA has defined GMP responsibilities for others. Exemption
from the GMP requirements does not exempt manufacturers of finished devices
from keeping complaint files ( 21
CFR 820.198) or from general requirements concerning records 21
CFR 820.180. Sterile devices are never exempted from GMP requirements. Medical
devices manufactured under an investigational device exemption (IDE) are not
exempt from design control requirements under 21 CFR 820.30. A device that normally
would be subject to GMP requirements may be exempt under the following conditions:
When FDA has issued an exemption order in response to a citizen's petition for
exemption;
When FDA, in the absence of a petition, has exempted the device and published the
exemption in the Federal Register;
When the device is exempted by FDA classification regulations published in the Federal
Register and codified in 21 CFR 862 to 892;
When the device is an intraocular lens (IOL) under an IDE and meets the requirements of
the IDE regulation for IOL's (except for design controls 21 CFR 820.30); and
Through a policy statement, FDA may decide not to apply GMP requirements to some types
of devices and processes although the devices may not have been exempted from GMP
requirements.
Manufacturers should be aware of the GMP exemption status of their devices. In
addition, manufacturers should keep on file records of any specific GMP exemption granted
to them by FDA. Upon request during a factory visit, the exemption records need to be
shown during normal business hours to the FDA investigator in order to verify that an
exemption has been granted.
Types of Establishments Exempt from GMP
Component Manufacturers
A "component" is defined by 21
CFR 820.3 (c) as "any raw material, substance, piece, part, software,
firmware, labeling, or assembly which is intended to be included as part of
the finished, packaged, and labeled device." Component manufacturers are
excluded from the QS regulation by 21
CFR 820.1 (a)(1). Current FDA policy is to rely upon the finished device
manufacturer to assure that components are acceptable for use. Component manufacturers
are not routinely scheduled for GMP inspections; however, FDA encourages them
to use the QS regulation as guidance for their quality system.
When finished device manufacturers produce components specifically for use
in medical devices they produce, whether in the same building or another location,
such production of components is considered part of the device manufacturing
operations, and the production should comply with the QS regulation as detailed
under Manufacturers of Accessories .
Types of Establishments Subject to the GMP
A remanufacturer as defined in 21
CFR 820.3 (w) is any person who processes, conditions, renovates, repackages,
restores, or does any other act to a finished device that changes the finished
device's performance or safety specifications or intended use. Remanufacturers
are considered manufacturers. As such, these manufacturers are subject to inspection
by FDA and shall meet the applicable requirements of the medical device QS regulation.
These manufacturers shall establish and implement quality systems to assure
the safety and effectiveness of the devices that are distributed. Such activities
include drafting of device master records, rebuilding per the device master
records, inspection and testing, calibration of measurement equipment, control
of components, updating of labeling, processing of complaints, and any other
GMP requirement applicable to the activities being performed.
Remanufacturers are also required to comply with the labeling requirements
of 21
CFR 801.1 (c). This labeling regulation requires that where the person or
manufacturer named on the label of the device is not the original manufacturer,
the name shall be qualified by an appropriate phrase which reveals the connection
that person has with the device, e.g., remanufactured by XYZ Company.
Section 520(b) of the FD&C Act and the IDE regulation (21 CFR 812.3 (b)) defines a custom device. Custom devices are exempt from certain statutory requirements. For example, manufacturers of custom devices are not required to comply with premarket approval requirements (Section 515) and are exempt from premarket notification requirements [Section 510(k)]. Custom devices are NOT exempt from the GMP requirements. Manufacturers of custom devices should comply with the GMP requirements while considering the flexibility allowed.
A person(s) that manufactures a finished device under the terms of a contract with another manufacturer is a contract manufacturer. The agreement between the manufacturers should be documented in a written contract. Contract manufacturers of finished devices shall comply with applicable requirements of the quality system and shall register their establishment with FDA. Depending on the circumstances, both the contractor and manufacturer may be held jointly responsible by FDA for the activities performed.
Contract laboratories that designs or test components or finished devices for a manufacturer according to the manufacturer's specifications are considered an extension of the manufacturer's quality system. These laboratories may provide services to a number of customers, many of which are not medical device manufacturers. These contract laboratories are not subject to routine GMP inspections. Through the conduct of purchasing assessment, the finished device manufacturer is responsible for assuring that equipment and procedures used by a lab are adequate and appropriate (21 CFR 820.50). However, an internal test laboratory, if part of a manufacturer that does testing for various facilities within the corporation, is subject to inspection when FDA GMP inspections are conducted at the individual manufacturing facilities. That is, the test laboratory is simply a part of a medical device manufacturer of which all device-related divisions shall comply with the QS regulation.
Repackagers, Relabelers, and Specification Developers
Repackaging and relabeling of a device and specification development are defined
as manufacturing in 21 CFR 820.3(o) and 21 CFR Part 807, Establishment
Registration and Device Listing for Manufacturers of Devices. Some definitions
from 21
CFR 807.3 (d) are reprinted below because they affect the applications of
the QS regulation.
(d) "Manufacture, preparation, propagating, compounding, assembly, or
processing" of a device means the making by chemical, physical, biological, or other
procedures of any article that meets the definition of a device in section 201(h) of the
Act. These terms include the following activities:
1. Repackaging or otherwise changing the container, wrapper, or labeling of any device
package in furtherance of the distribution of the device from the original place of
manufacture to the person who makes final delivery or sale to the ultimate consumer;
2.Distribution of domestic or imported devices; or
3. Initiation of specifications for devices that are manufactured by a second party for
subsequent commercial distribution by the person initiating specifications".
As defined above, repackaging and relabeling are manufacturing operations.
Further, a repacker, repackager or relabeler is a manufacturer per 21
CFR 820.3 (o) and subject to the applicable requirements of the QS regulation.
Individuals are repackers or relabelers if they:
Individuals are not considered repackers or relabelers or a manufacturer for purposes
of applying the QS regulation if they pack only previously packaged and labeled individual
devices into packages for the convenience of the user. (Note that this activity is
essentially the same as a drug store employee placing packaged items into a bag labeled
with the name of the drug store.)
A distributor who only adds a label bearing their name and address is exempt from the
GMP requirements. A manufacturer simply affixing a sticker label bearing the distributor's
name and address would not require record keeping demonstrating compliance with labeling
controls requirements.
Specification developers provide specifications to contract manufacturers, who produce
devices to meet the specifications. The contract manufacturer may package and label the
device, or the finished device may be shipped to the specification developer for packaging
and labeling.
Specification developers are manufacturers and are subject to the GMP requirements
that apply to the activities they conduct, such as various design controls including
correct transfer of the design information to a contract manufacturer [21
CFR 820.30 (h)]. This activity, in turn, requires an adequate device master
record (21
CFR 820.181) and adequate document change control [21
CFR 820.40 (b)]. Further, if the product carries the specification developer's
label, the developer is responsible for maintaining a complaint file and processing
complaints, plus maintaining the device specifications and other appropriate
documents in the device master record.
When finished device manufacturers produce components specifically for use in medical
devices they produce, whether in the same building or another location, such production of
components is considered part of the device manufacturing operations, and the production
should comply with the QS regulation.
Accessory devices are discussed in 21 CFR Part 807, Establishment Registration and Device Listing for Manufacturers of Devices [21 CFR 807.20 (a)(5)]. These devices, such as hemodialysis tubing or major diagnostic xray components, that are packaged, labeled, and distributed separately to a hospital, physician, etc., for healthrelated purposes are sometimes inappropriately referred to as components. However, FDA considers them finished devices because they are suitable for use or capable of functioning and are distributed for health-related purposes; and the QS regulation applies to their manufacture. Similarly, a device or component including software that is sold as an addition to a finished medical device to augment or supplement its performance is also termed an accessory. An accessory to a medical device is considered a finished device and, therefore, is subject to the QS regulation.
Initial Distributors of Imported Devices
An initial importer (or initial distributor) takes first title to the devices imported into the U.S. and further distributes the product. With regards to the GMP, this initial distributor is responsible for maintaining complaint files and general record keeping requirements. A procedure shall be established and maintained for receiving, reviewing, and evaluating complaints. All complaints, including oral complaints, are to be processed in a uniform and timely manner. These complaints shall be evaluated to determine whether or not they require reporting to FDA under Medical Device Reporting, 21 CFR Part 804 or 803. The initial distributor is also required to evaluate all complaints to determine whether an investigation is necessary, as well as complying with all other requirements in 21 CFR 820.198, Complaint Files.
Additional Quality System Information