FDA
Home Page | CDRH Home Page | Search
| CDRH
A-Z Index | Contact CDRH
|
Thermal Endometrial Ablation DevicesSUBMISSION GUIDANCE FOR AN IDE |
This guidance was written prior to the February 27, 1997 implementation of FDA's Good Guidance Practices, GGP's. It does not create or confer rights for or on any person and does not operate to bind FDA or the public. An alternative approach may be used if such approach satisfies the requirements of the applicable statute, regulations, or both. This guidance will be updated in the next revision to include the standard elements of GGP's.
FINAL: March 14, 1996
Developed by:
Obstetrics and Gynecology Devices Branch
Office of Device Evaluation, HFZ-470
9200 Corporate Blvd
Rockville, MD 20850
(301) 594-1180
Summary of all prior investigations
Feasibility Study - Effectiveness
Safety and Effectiveness Study
This guidance is intended to identify the elements that the Office of Device Evaluation (ODE) would expect to see in an Investigational Devices Exemptions (IDE) application for a Thermal Endometrial Ablation Device. This group of devices includes new types of devices that are intended to ablate the endometrium using conductive heating, RF energy, microwave energy, cryosurgery, etc., and where the user has no direct control over the course of the procedure once it has begun (i.e., no feedback). It is important to understand that certain technologies may not require all of the information contained herein, whereas other technologies may require additional studies beyond the scope of this guidance document.
For general information about how to submit an IDE application, contact the Center for Devices and Radiological Health's (CDRH) Division of Small Manufacturers Assistance (DSMA) at (800) 638-2041 or (301) 443-6597. FDA welcomes comments on this draft guidance document and will consider all scientifically valid alternatives to the preclinical and clinical requirements stated within. It is also highly recommended that the sponsor of a new investigation contact the Obstetrics and Gynecology Devices Branch (OGDB) within the Office of Device Evaluation (ODE) prior to submission of an original IDE application, at (301) 594-1180.
A thorough understanding of device design is essential not only in evaluating the functional characteristics of the device but also in evaluating the clinical application of the device. Provide a brief overview of the principles of operation of the device. This should include:
Provide a complete list of all patient-contacting materials, and where relevant, provide a discussion as to why a given material was chosen for a particular function. If any patient-contacting material contains a color pigment, please provide the following information: chemical composition, color index number, and color additive listing (from 21 CFR 73).
Provide either biocompatibility testing (performed on the finished device), or certification that identical materials are used in a legally marketed device with a similar intended use. Samples for biocompatibility testing should be prepared in such a way so as to reflect the actual conditions of use (e.g., if the material will be heated to 90ºC during use, it should be heated to this temperature prior to testing). Tests should be conducted in conformance with Good Laboratory Practices (GLP) in accordance with 21 CFR 58.
For additional information on biocompatibility, please refer to the Blue Book Memorandum "Use of International Standard ISO-10993, `Biological Evaluation of Medical Devices Part 1: Evaluation and Testing'", available from DSMA.
Please provide documentation describing the software development lifecycle and risk management activities. This should include:
Additional guidance on software documentation can be found in "Reviewer Guidance for Computer Controlled Medical Devices Undergoing 510(k) Review", available from DSMA.
Provide either:
Provide either:
Provide system-level testing demonstrating that the device performs as designed (e.g., it heats to the desired temperature, thermocouples accurately record temperature, etc.). Include:
Provide samples of all device labeling. The labeling must include the following information (21 CFR 812.5):
Note: 21 CFR 812.5 (b) stipulates that the labeling of an investigational device shall not bear any statement that is false or misleading in any particular and shall not represent that the device is safe or effective for the purposes for which it is being investigated.
Provide a description of the methods, facilities, and controls used for the manufacture, processing, packaging, and storage of the device, in sufficient detail so that a person generally familiar with Good Manufacturing Practices can make a knowledgeable judgement about the quality control used in the manufacture of the device.
Devices that are intended to contact or enter sterile tissue or a normally sterile body cavity, such as the uterus, require sterilization before use.
Provide detailed instructions for reprocessing (cleaning and sterilizing) the device, including instructions necessary for any assembly/disassembly. Also, include a warning that the device should be thoroughly cleaned and sterilized according to validated infection control procedures before use/reuse.
For important additional information on reprocessing, please refer to the draft "Labeling Reusable Medical Devices Reprocessing in Health Care Facilities: FDA Reviewer Guidance" (March 1995). A copy of this guidance may be obtained from DSMA.
Specify whether or not the device is to be sold to the patient during the clinical study. If so, explain why this does not constitute commercialization.
Provide either:
Provide a complete description of all in vitro (e.g., extirpated uteri) and animal testing. This should include:
Note: Adequate data from in vitro and animal testing, demonstrating that the device is relatively safe and that it functions as intended, must be submitted before approval for clinical studies in humans will be granted.
Provide the following information about all prior clinical investigations involving the device:
The purpose of the Feasibility Safety Study is to demonstrate the safety of the device in a limited number of women already scheduled for a hysterectomy. This involves monitoring temperatures within the uterine wall and on the serosal surface during the ablation procedure, and conducting a histological examination of the uterus once it is removed.
Significant changes to the temperature/time parameters or to the system design may require revalidation via an additional feasibility study.
Note: If the safety and effectiveness study will include women who have had previous uterine surgery (e.g., caesarean section, myomectomy), then the safety study should include sufficient numbers of both women who have had pervious uterine surgery and those who have not.
(1) pelvic examination
(2) Pap smear
(3) pregnancy test
(4) endometrial biopsy
All patients should receive either hormonal pre-treatment to thin the endometrium, or an immediate pre-procedure D&C. All subjects should receive the same pre-treatment regimen.
(1) Provide the procedures used for the ablation procedure. Ablation parameters should be exactly the same as those to be used in the safety and effectiveness study. If they are not, provide a detailed discussion of why results from the feasibility study will still be relevant.
(2) The following data should be collected:
(a) temperature measurements during the ablation procedure, including the location of the thermocouples or other temperature measuring devices.
(b) histology, including both gross and microscopic.
(c) depth of penetration of thermal injury.
(3) Provide a complete discussion of any adverse events.
Patients should be followed-up according to the standard of care for hysterectomy.
Provide a complete description of all potential risks to the patient.
Provide copies of the informed consent forms that will be used during the study. You may wish to consult the "Investigational Device Exemption Manual", available from DSMA, for additional guidance on informed consent.
The purpose of this study is to establish tentative effectiveness rates for use in sample size calculations, and to support beginning a full-scale safety and effectiveness study. This is a single arm feasibility study in at least 20 women who are candidates for endometrial ablation.
Note: Previously collected data from studies conducted at foreign sites may be accepted in lieu of this study. In this case, the following information should be provided:
The study sponsor must determine the appropriate hypothesis and study endpoints. Primary study endpoints should be quantitative in nature (e.g., reduction in uterine bleeding, elimination of uterine bleeding (amenorrhea), increase in hematocrit). However, studies should also include a secondary quality-of-life endpoint.
Six months of follow-up data from this "Feasibility - Effectiveness" study should be provided to FDA prior to submission of an IDE for the safety and efficacy study. In addition, all patients should be followed for a total of 2 years (i.e., 6 months pre-submission plus 18 months post-submission). Once all data are collected, a final report should be submitted to FDA.
(a) pelvic examination
(b) Pap smear
(c) pregnancy test
(d) hematocrit
(e) diagnostic hysteroscopy with endometrial biopsy
If patient blood loss is to be used as a study endpoint, baseline data should be collected for three months prior to the ablation procedure.
Patients should receive either hormonal pre-treatment to thin the endometrium, or an immediate pre-procedure D&C. All subjects should receive the same pre-treatment regimen.
Provide the following information for each patient:
(1) The procedures used for the ablation procedure. Ablation parameters should be exactly the same as those to be used in the safety and effectiveness study. If they are not, provide a detailed discussion of why the results will still be relevant.
(2) Adverse events
Patients should receive follow-up examinations at the following intervals: 2 weeks, 3 months, and 6 months. The study sponsor should account for all patients lost to follow-up.
Provide a complete description of all risks to the patient.
Provide copies of the informed consent forms that will be used during the study. You may wish to consult the "Investigational Device Exemption Manual", available from DSMA, for additional guidance on informed consent.
The purpose of this study is to obtain the safety and effectiveness data necessary to support a PMA.
(1) The study should compare the investigational device to a legally marketed device intended for endometrial ablation (e.g., electrosurgical resection, rollerball ablation, or laser ablation). Control arms such as sham or D&C are not acceptable.
(2) Describe the methods used to ensure uniformity of the control procedure from investigator to investigator. A single control procedure should be chosen, and investigators must be of recognized competence in performing the procedure.
(3) Multi-Center
The study should be conducted at several centers, with at least 10 patients from each procedure at each center.
(4) Describe the methods that will be used to minimize bias, including:
(a) randomization of patient selection process
(b) procedures intended to prevent "shopping" for treatment
The study sponsor must determine the appropriate hypothesis and study endpoints. Primary study endpoints should be quantitative in nature (e.g., reduction in uterine bleeding, elimination of uterine bleeding (amenorrhea), increase in hematocrit). However, studies should also include a secondary quality-of-life endpoint, based on a scaled questionnaire. In addition, the sponsor should also collect data on the amenorrhea rate and the need for future surgical interventions.
(1) provide appropriate statistical justification
(2) adequate numbers at each site - see C.1.a (2) above.
12 months of follow-up data will be required prior to PMA approval. However, the PMA may be submitted once 6 months of data has been obtained for all subjects. In addition, all patients should be followed for a total of 3 years (i.e., 1 year pre-approval plus 2 years in a post-market setting). Data collection in the post-approval setting should focus on the need for repeat ablations or hysterectomy - qualitative blood loss data need not be collected. Once all data are collected, a final report should be submitted to FDA as a PMA supplement.
(1) pelvic examination
(2) pap smear
(3) pregnancy test
(4) hematocrit
(5) endometrial biopsy
(6) uterine evaluation by one of the following: diagnostic hysteroscopy, transvaginal ultrasound, or hysterosalpingogram (HSG)
If patient blood loss is to be used as a study endpoint, baseline data should be collected for three months prior to the ablation procedure.
Patients in both study arms should receive either hormonal pre-treatment to thin the endometrium, or an immediate pre-procedure D&C. All subjects should receive the same pre-treatment regimen.
Patients should receive follow-up examinations at the following intervals:
(1) pelvic exam
(2) adverse event report
(1) pelvic exam
(2) adverse event report
(3) hematocrit
(4) evaluation of diaries and Quality of Life
(1) pelvic exam
(2) adverse event report
(3) hematocrit
(4) evaluation of diaries and Quality of Life
(1) adverse event report
(2) evaluation of diaries and Quality of Life
Provide a complete description of all risks to the patient.
Provide copies of the informed consent forms that will be used during the study. You may wish to consult the "Investigational Device Exemption Manual", available from DSMA, for additional guidance on informed consent.
The following analyses should be included as part of the PMA.
The following related documents are available from the Center for Devices and Radiological Health's (CDRH) Division of Small Manufacturers Assistance (DSMA) at (800) 638-2041 or (301) 443-6597.
Blue Book Memorandum "Use of International Standard ISO-10993, `Biological Evaluation of Medical Devices Part 1: Evaluation and Testing'"
Reviewer Guidance for Computer Controlled Medical Devices Undergoing 510(k) Review
Labeling Reusable Medical Devices Reprocessing in Health Care Facilities: FDA Reviewer Guidance (March 1995)
Investigational Device Exemption Manual
Brill, A.I. (1995) What is the role of hysteroscopy in the management of abnormal uterine bleeding? Clinical Obstetrics and Gynecology, 38(2), pp.319-345.
Chimbira, T.H. & A.C. Turnbull (1980) Relation between measured menstrual blood loss and patient's subjective assessment of loss, duration of bleeding, number of sanitary towels used, uterine weight and endometrial surface area, British Journal of Obstetrics and Gynaecology, 87, pp.603-609.
Copperman, A.B., DeCherney, A.H. & D.L. Olive (1993) A case of endometrial cancer following endometrial ablation for dysfunctional uterine bleeding, Obstetrics and Gynecology, 82, pp.640-642.
DeCherney, A.H., Diamond, M.P., Lavy, G. & M.L. Polan (1987) Endometrial ablation for intractable uterine bleeding: hysteroscopic resection, Obstetrics and Gynecology, 70, pp.668-670.
Deeny, M. and J.A. Davis (1994) Amount of menstrual blood loss in women referred for endometrial ablation, European Journal of Obstetrics and Gynecology, 57, pp.179-180.
Eijkeren, M.A. et al (1986) The alkaline hematin method for measuring menstrual blood loss - a modification and its clinical use in menorrhagia, European Journal of Obstetrics Gynecology and Reproductive Biology, 22, pp.345-351.
Erian, J. (1994) Endometrial ablation in the treatment of menorrhagia, British Journal of Obstetrics and Gynaecology, 101 (Supplement 11), pp.19-22.
Fraser, I. S. (1994) Menorrhagia - a pragmatic approach to the understanding of causes and the need for investigations, British Journal of Obstetrics and Gynaecology, 101 (Supplement 11), pp.3-7.
Fraser, I.S., Angsuwathana, S., Mahmoud, F. & S. Yezerski (1993) Short and medium term outcomes after rollerball endometrial ablation for menorrhagia, Medical Journal of Australia, 158, pp.454-457.
Fraser, I.S., McCarron, G. & R. Markham (1984) A preliminary study of factors influencing perception of menstrual blood loss volume, Obstetrics and Gynecology, 149, pp.788-793.
Garry, R., Shelley-Jones, D., Mooney, P. & G. Phillips (1995) Six hundred endometrial laser ablations, Obstetrics and Gynecology, 85, pp.24-29.
Garry, R. (1995) Good practice with endometrial ablation, Obstetrics and Gynecology, 85, pp.144-151.
Hallberg, L., Högdahl, A.-M., Nilsson, L, & G. Rybo (1966) Menstrual blood loss - a population study, Acta Obstet. et Gynecol. Scandinav., 45, 320-351.
Janssen, C.A.H., Scholtenm P.C. & A.P.M Heintz (1995) A simple visual assessment technique to discriminate between menorrhagia and normal menstrual blood loss, Obstetrics and Gynecology, 85, pp.977-982.
Jenkinson, C., Peto, V. & A. Coulter (1994) Measuring change over time: a comparison of results from global single item of health status and the multi-dimensional SF-36 health status survey questionnaire in patients presenting with menorrhagia, Quality of Life Research, 3, pp.317-321.
Lalonde, A. (1994) Evaluation of surgical options in menorrhagia, British Journal of Obstetrics and Gynaecology, 101 (Supplement 11), pp.8-14.
Lefler, H.T. & C.F. Lefler (1992) Ablation of the endometrium: three-year follow up for perimenstrual symptoms, Journal of Reproductive Medicine, 37(2), pp.147-150.
Lewis, B.V. (1994) Guidelines for endometrial ablation, British Journal of Obstetrics and Gynaecology, 101, pp.470-473.
Loffer, F. D. (1995) Good practice with endometrial ablation (letter), Obstetrics and Gynecology, 86(5).
Magos, A., Baumann, R., Lockwood, G.M. & A.C. Turnball (1991) Experience with the first 250 endometrial resections for menorrhagia, Lancet, 377, pp.1074-1078.
Newton, J., Barnard, G. & W. Collins (1977) A rapid method for measuring menstrual blood loss using automatic extraction, Contraception, 16(3), pp.269-282.
Onbargi, L.C., Hayden, R., Valle, R.F. & G.D. Priore (1993) Effects of power and electrical current density variations in an in vitro endometrial ablation model, Obstetrics and Gynecology, 82, pp.912-918.
Pinion, S.B. et al (1994) Randomised trial of hysterectomy, endometrial laser ablation, and transcervical endometrial resection for dysfunctional uterine bleeding, British Medical Journal, 309(6960), pp.979-983.
Ramey, J.W. et al. (1994) The process of carcinogenesis for endometrial adenocarcinoma could be short: development of a malignancy after endometrial ablation, Obstetrics and Gynecology, 170, pp.1370-1371.
Ruta, D.A. et al. (1995) Assessment of patients with menorrhagia: how valid is a structured clinical history as a measure of health status? Quality of Life Research, 4, pp.33-40.
Scottish Hysteroscopy Audit Group (1995) A Scottish audit of hysteroscopic surgery for menorrhagia: complications and follow up, British Journal of Obstetrics and Gynaecology, 102, pp.249-254.
Shaw, R.W. (1994) Assessment of medical treatments for menorrhagia, British Journal of Obstetrics and Gynaecology, 101 (Supplement 11), pp.15-18.
Sørensen, S.S., Andersen, L.F. & G. Lose (1994) Endometriosis by implantation: a complication of endometrial ablation, Lancet, 343, p.1226.
Sporrong, T., Rybo, G. & L.A. Mattsson (1992) An objective and subjective assessment of uterine blood loss in postmenopausal women on hormone replacement therapy, British Journal of Obstetrics and Gynaecology, 99, pp.399-401.
Sutton, C.J.G. (1994) Thinning the endometrium prior to ablation: is it worthwhile? British Journal of Obstetrics and Gynaecology, 101 (Supplement 10), pp.10-12.
Townsend, D.E. et al. (1993) Post-ablation-tubal sterilization syndrome, Obstetrics and Gynecology, 82, pp.422-424.
Treatment of dysfunctional uterine bleeding (letters) (1995) British Medical Journal, 310(9982), pp.801-804.
Valle, R.F. (1993) Endometrial ablation of dysfunctional uterine bleeding: role of GnRH agonists, International Journal of Gynecology and Obstetrics, 41, pp.3-15.
Velevil, P. et al (1995) Rate of hospitalization for gynecologic disorders among reproductive-age women in the United States, Obstetrics and Gynecology, 86(5), pp.764-769.
Wathen, P.I., Henderson, M.C. & C.A. Witz. (1995) Abnormal uterine bleeding, Medical Clinics of North America, 79(2), pp.329-344.
Uploaded on October 26, 1998
CDRH Home Page | CDRH A-Z Index | Contact CDRH | Accessibility | Disclaimer
FDA Home Page | Search FDA Site | FDA A-Z Index | Contact FDA | HHS Home Page
Center for Devices and Radiological Health / CDRH