Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA

See Related Info

PDF Printer Version

Document issued on: February 5, 2003

 

This document supersedes Guidance on Review Criteria for Assessment of Antimicrobial
Susceptibility Devices; Draft, March 8, 2000

 

 

 

 

CDRH Logo

U.S. Department of Health and Human Services
Food and Drug Administration
Center for Devices and Radiological Health

Division of Microbiology Devices
Office of In Vitro Diagnostic Device Evaluation and Safety


Preface

Public Comment:

Comments and suggestions may be submitted at any time for Agency consideration to Dockets Management Branch, Division of Management Systems and Policy, Office of Human Resources and Management Services, Food and Drug Administration, 5630 Fishers Lane, Room 1061, (HFA-305), Rockville, MD, 20852. When submitting comments, please refer to Docket No. 00D-0109. Comments may not be acted upon by the Agency until the document is next revised or updated.

For questions regarding the use or interpretation of this guidance contact Freddie Poole at (301) 594-2096 or by email FMP@cdrh.fda.gov.

Additional Copies:

Additional copies are available from the Internet at: http://www.fda.gov/cdrh/ode/631.pdf, or you may send a fax request to (301) 443-8818 to receive a hard copy of the document, or send an e-mail request to GWA@CDRH.FDA.GOV to request hard or electronic copy. Please use the document number (631) to identify the guidance you are requesting.

Table of Contents

I. Introduction

II. Background

III. The Least Burdensome Approach

IV. The Content and Format of an Abbreviated 510(k) Submission

V. Scope

VI. Device Description

VII. Risks to Health

VIII. Device History

IX. Study Design

A. Reference Method

B. New Device

C. Organism Selection

1. Fresh clinical organisms

2. Clinical stock organisms

3. Challenge organisms

D. Quality Control

1. Selection of Quality Control Organisms

2. Inoculum density check

3. Purity check

E. Reproducibility

X. Data Presentation

A. Comparative Performance Data

1. Clinical – fresh and stock

2. Challenge organisms

3. Challenge plus clinical organisms

B. Quality Control

C. Reproducibility

XI. Evaluating the Results of your Study

A. Fresh, Stock, and Challenge Organisms

B. Quality Control

C. Reproducibility

XII. Labeling

A. Intended Use Statement

B. Reagents

C. Reporting of Results

D. Performance Characteristics

E. Quality Control

F. Limitations

XIII. Removing Certain Labeling Limitations from Legally Marketed AST Devices

A. Performance

B. Insufficient Resistant Strains

C. Reproducibility

D. Quality Control

XIV. QSR Considerations

A. Lot to Lot Reproducibility

B. Stability

XV. Glossary

XVI. References

TABLE 1: Recommendations for Antimicrobial Susceptibility Devices

TABLE 2: Recommendations for the Removal of Limitations from Antimicrobial Susceptibility Devices

TABLE 3: Presentation of Summary Data for Both Challenge and Clinical Data

TABLE 4: Example of Reporting Format for Quality Control Data

TABLE 5: Sample Table Format for Device Performance

TABLE 5A: Sample Table Format for Device Performance, continued

TABLE 6: Presentation of Reproducibility Results by Organism

TABLE 6A: Presentation of Reproducibility Results by Organism and Site

TABLE 6B: Presentation of Reproducibility Results by Organism, Pooled Across Sites

TABLE 7: Report Format for Inoculum Density

TABLE 8: Number of VMJ Discrepancies as a Function of the Number of Resistant Organisms Tested

TABLE 9: Essential Agreement as Function of the Number of Evaluable Organisms Tested

Class II Special Controls Guidance
Document: Antimicrobial Susceptibility Test
(AST) Systems; Guidance for Industry and
FDA

This document is intended to provide guidance. It represents the Agency’s current thinking on this topic. It does not create or confer any rights for or on any person and does not operate to bind the Food and Drug Administration (FDA) or the public. An alternative approach may be used if such approach satisfies the requirements of the applicable statute and regulations.

I. Introduction

This guidance document was developed as a special control guidance to support the reclassification of the antimicrobial susceptibility test (AST) system, when the device is a system employing short-term incubation (less than 16 hours) from class III into class II (special controls). The device, as proposed, is intended to determine the in vitro susceptibility of bacterial pathogens from clinical specimens. This guidance is issued in conjunction with a Federal Register notice announcing the reclassification of the automated short-term incubation cycle AST system.

Following the effective date of this final reclassification rule, any firm submitting a 510(k) premarket notification for aan automated short-term incubation cycle AST system will need to address the issues covered in the special control guidance. However, the firm need only show that its device meets the recommendations of the guidance or in some other way provides equivalent assurances of safety and effectiveness.

II. Background

FDA believes that special controls, when combined with the general controls, will be sufficient to provide reasonable assurance of the safety and effectiveness of the automated short-term incubation cycle AST system. Thus, a manufacturer who intends to market a device of this generic type should (1) conform to the general controls of the Federal Food, Drug & Cosmetic Act (the Act), including the premarket notification requirements described in 21 CFR 807 Subpart E, (2) address the specific risks to health associated with the automated short-term incubation cycle AST system identified in this guidance and, (3) obtain a substantial equivalence determination from FDA prior to marketing the device (see also 21 CFR 807.85).

This special control guidance document identifies the classification regulations and product codes for the automated short-term incubation cycle AST system (Refer to Section V – Scope). In addition, other sections of this special control guidance document list the risks to health identified by FDA and describe measures that, if followed by manufacturers and combined with the general controls, will generally address the risks associated with these automated short-term incubation cycle AST systems and lead to a timely premarket notification [510(k)] review and clearance. This document supplements other FDA documents regarding the specific content requirements of a premarket notification submission. You should also refer to 21 CFR 807.87 and other FDA documents on this topic, such as the 510(k) Manual - Premarket Notification: 510(k) - Regulatory Requirements for Medical Devices,    http://www.fda.gov/cdrh/manual/510kprt1.html.

Under "The New 510(k) Paradigm - Alternate Approaches to Demonstrating Substantial Equivalence in Premarket Notifications; Final Guidance1," a manufacturer may submit a Traditional 510(k) or has the option of submitting either an Abbreviated 510(k) or a Special 510(k). FDA believes an Abbreviated 510(k) provides the least burdensome means of demonstrating substantial equivalence for a new device, particularly once a special controls guidance document has been issued. Manufacturers considering modifications to their own cleared devices may lessen the regulatory burden by submitting a Special 510(k).

_____________________
1 http://www.fda.gov/cdrh/ode/parad510.html

III. The Least Burdensome Approach

The issues identified in this guidance document represent those that we believe need to be addressed before your device can be marketed. In developing the guidance, we carefully considered the relevant statutory criteria for Agency decision-making. We also considered the burden that may be incurred in your attempt to comply with the guidance and address the issues we have identified. We believe that we have considered the least burdensome approach to resolving the issues presented in the guidance document. If, however, you believe there is a less burdensome way to address the issues, you should follow the procedures outlined in the A Suggested Approach to Resolving Least Burdensome Issues document. It is available on our Center web page at: 
http://www.fda.gov/cdrh/modact/leastburdensome.html

IV. The Content and Format of an Abbreviated 510(k) Submission

An Abbreviated 510(k) submission must include the required elements identified in 21 CFR 807.87, including the proposed labeling for the device sufficient to describe the device, its intended use, and the directions for its use. In an Abbreviated 510(k), FDA may consider the contents of a summary report to be appropriate supporting data within the meaning of 21 CFR 807.87(f) or (g); therefore, we recommend that you include a summary report. The report should describe how this special control guidance document was used during the device development and testing and should briefly describe the methods or tests used and a summary of the test data or description of the acceptance criteria applied to address the risks identified in this guidance document, as well as any additional risks specific to your device. This section suggests information to fulfill some of the requirements of 807.87 as well as some other items that we recommend you include in an Abbreviated 510(k).

Coversheet

The coversheet should prominently identify the submission as an Abbreviated 510(k) and cite the title of this class II special controls guidance document.

Proposed labeling

Proposed labeling should be sufficient to describe the device, its intended use, and the directions for its use. (Refer to Section XII for specific information that should be included in the labeling for devices of the types covered by this document.)

Summary report

We recommend that the summary report contain:

_____________________
2 Refer to http://www.fda.gov/cdrh/ode/indicate.html for the recommended format.
3 If FDA makes a substantial equivalence determination based on acceptance criteria, the subject device should be tested and shown to meet these acceptance criteria before being introduced into interstate commerce. If the finished device does not meet the acceptance criteria and, thus, differs from the device described in the cleared 510(k), FDA recommends that submitters apply the same criteria used to assess modifications to legally marketed devices (21 CFR 807.81(a)(3)) to determine whether marketing of the finished device requires clearance of a new 510(k).  .
4 See Required Elements for a Declaration of Conformity to a Recognized Standard (Screening Checklist for All Premarket Notification [510(K)] Submissions), http://www.fda.gov/cdrh/ode/reqrecstand.html

If it is not clear how you have addressed the risks identified by FDA or additional risks identified through your risk analysis, we may request additional information about aspects of the device’s performance characteristics. We may also request additional information if we need it to assess the adequacy of your acceptance criteria. (Under 21 CFR 807.87(l), we may request any additional information that is necessary to reach a determination regarding substantial equivalence.)

As an alternative to submitting an Abbreviated 510(k), you can submit a Traditional 510(k) that provides all of the information and data required under 21 CFR 807.87 and described in this guidance. A Traditional 510(k) should include all of your methods, data, acceptance criteria, and conclusions. Manufacturers considering modifications to their own cleared devices should consider submitting Special 510(k)s.

The general discussion above applies to any device subject to a special controls guidance document. The following is a specific discussion of how you should apply this special controls guidance document to a premarket notification for an automated short-term incubation cycle AST system.

V. Scope

The scope of this document is limited to the following device as described in 21 CFR 866.1645 Fully automated short-term incubation cycle antimicrobial susceptibility system (product code: LON).

The classification identification below identifies the device as it existed at the time of reclassification:

An antimicrobial susceptibility test system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term incubation (less than 16 hours) are used to determine the antimicrobial agent of choice to treat bacterial diseases.

This document does not apply to devices intended for testing anti-mycobacterial, anti-viral, or anti-fungal agents or devices intended for testing the susceptibility of fastidious organisms for which there is no NCCLS standard reference method for testing.

Devices classified in section 21 CFR 866.1640, Antimicrobial susceptibility test powder (product codes shown below) are not subject to this special control guidance. However, information in this document may be useful to manufacturers of these devices.

This document does not address antimicrobial disks for the disk diffusion method classified in section 21 CFR 866.1620. These devices are addressed in the guidance, "Review Criteria for Assessment of Antimicrobial Susceptibility Test Discs," http://www.fda.gov/cdrh/ode/testdisc.pdf.

VI. Device Description

You should identify your device by regulation and product code and a legally marketed predicate device.

In order to help FDA quickly view all the aspects of your device compared with the predicate, you should include a table that outlines the similarities and differences between the predicate and your device.

VII. Risks to Health

In the table below, FDA has identified the risk to health generally associated with the use of the automated short-term incubation cycle AST system addressed in this document. The measures recommended to mitigate the identified risk are given in this guidance document, as shown in the table below. You should also conduct a risk analysis, prior to submitting your premarket notification, to identify any other risks specific to your device. The premarket notification should describe the risk analysis method. If you elect to use an alternative approach to address the risk identified in this guidance document, or have identified risks additional to those in the guidance, you should provide sufficient detail to support the approach you have used to address that risk.

Identified risk Recommended mitigation measures

administration of an inappropriate
antimicrobial agent to a patient

Sections IX, X, XI

VIII. Device History

This guidance document ensures well-standardized, reliable, and reproducible performance evaluation for AST devices. Clinically, results from AST devices are useful for therapeutic guidance whenever the susceptibility of a bacterial pathogen may be unpredictable or when the infecting organism belongs to a species that may be resistant to antimicrobial agents of choice. Additionally, susceptibility testing is useful for monitoring development of new or emerging resistance to antimicrobial agents.

A determination of substantial equivalence to a legally marketed predicate device is based on intended use, design, energy used or delivered, materials, performance, safety, effectiveness, labeling, and other applicable characteristics. FDA believes performance of this type device is best established by comparison to the NCCLS standard reference methods (Ref. 1, 2) for each antimicrobial agent.

Laboratory procedures used for determining susceptibility of bacteria to antimicrobial agents have been developed and standardized over the past five decades. Historically, there have been two general procedures applied to susceptibility testing, i.e., dilution and diffusion. Other manual testing methods are based on modifications and refinements of older techniques such as gradient diffusion. Voluntary consensus standards on methodology and interpretive categories were implemented for susceptibility testing results that are antimicrobial agent, organism, or methodology dependent. NCCLS is the major organization in the United States that establishes voluntary standards and guidelines for standardizing and maintaining performance of laboratory susceptibility tests. A system has been established for continual assessment and upgrading of recommendations and addition of test criteria for new antimicrobial agents and older agents particularly when emerging resistance is recognized. A separate subcommittee was established in 1986 to standardize methods (Ref. 1, 2) for developing in vitro susceptibility testing criteria. These methods are also used by the pharmaceutical industry for developing new antimicrobial agents.

The NCCLS standard reference methods use 16-24 hours incubation for aerobic bacteria and 48 hours for anaerobic bacteria. Because shorter incubation times may provide clinical advantages, a number of manufacturers have developed automated procedures designed to generate results more rapidly, generally by the use of shortened incubation times (<16 hours). The results of reference overnight (16-24 hours of incubation) tests are accepted as standards for evaluating methods with a shortened incubation for the following reasons:

NCCLS has an Approved Standard M7 "Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically" (Ref. 1) that recommends a reference method for non-fastidious organisms. Other organisms that will not grow satisfactorily in (or on) unsupplemented Mueller-Hinton medium within 24 hours are considered fastidious organisms and may be included in NCCLS approved standards but generally with a different medium recommended for testing. If the FDA approved pharmaceutical antimicrobial agent package insert includes fastidious organisms (e.g., Streptococci, Haemophilus) with interpretive criteria and there is an NCCLS standard methodology, the recommendations for performance assessment are similar, but the numbers necessary for review may vary. See Table 1 for recommendations.

A susceptibility result may suggest that an uncomplicated bacterial infection can be effectively treated if AST device results indicate that the bacterial isolate is susceptible to the antimicrobial agent selected. The inability of a new device to produce a susceptible result for an organism that is susceptible to an antimicrobial agent by the reference method is considered a "major discrepancy." In this case, if a new device yields a resistant result for an organism, the antimicrobial agent may not be made available for treatment when in fact it could be an effective choice. Such major discrepancies can lead to utilization of broad-spectrum agents and needlessly accentuate the pressure for selection of resistant flora. Conversely, the inability to detect resistance is assessed by the "very major discrepancy rate," since therapy with that antimicrobial agent may lead to treatment failure, particularly for serious infections or altered host conditions. Accurate detection of resistance is important for clinical effectiveness and for monitoring emergence of resistance in the community.

Resistance to antimicrobial agents can generally be classified into four basic mechanisms:

The time needed for expression of resistance varies with different combinations of antimicrobial agents and organisms that have different mechanisms of resistance. The delay of expression of resistance can range from one to many hours. Studies comparing results of shorter incubation test results with conventional 16 to 24 hour incubation methods have documented the difficulties of detecting delayed resistance expression. Manufacturers of devices with shortened incubation times have adopted a variety of strategies to bring these results as close to conformity as possible when compared with results using the NCCLS standard reference methods. Examples of these strategies include:

Other devices can detect resistance by the presence or absence of a genotype associated with in vitro resistance.

IX. Study Design

Table 1 outlines, in tabular form, FDA’s recommendations for the number of sites, and type and numbers of organisms for testing.

Generally, FDA recommends that you establish the performance characteristics of your AST device by agreement with the NCCLS standard reference method for each antimicrobial agent and the organisms intended for testing. Because variations in test procedures can affect performance, we believe you should conduct agreement studies on all of the procedural options included in the directions for use section of the package insert. Such procedural options include, but are not limited to, inoculation preparation methods and reading of results, for example:

You should also address all possible combinations of these procedural options.

You should have a testing protocol describing testing procedures for both the reference method and new device. The protocols should include the exact procedures to follow for the reference and new device.

We recommend that you include your testing protocol in your 510(k). The protocol should describe your study design and contain the type of quality control recommended and the procedures for the reference and test method. The procedures should include:

For a valid comparison, FDA believes you should not deviate from the NCCLS standard reference method procedure. As stated above, you should include testing procedures for the new device with all procedural options, or possible combinations of these options that are included in the instructions of the package insert. This is especially important for certain organism-antimicrobial agent combinations that are affected by variations in inoculum and have growth patterns that may be interpreted differently when read visually or automatically. Your study design should include options for different methods of inoculation or additional dilutions of the inoculum suspension for certain groups of organisms (e.g., Proteus sp.), if any options are in the instructions for use.

Where appropriate to the device design and instructions for use, performance data using alternate methods of reading and/or inoculation procedures should include test results for all challenge, quality control, and reproducibility studies.

A.Reference Method

The reference method plates should contain two-fold dilutions of the antimicrobial agent for which FDA clearance is sought. The selection of dilutions should include the FDA and NCCLS interpretive standards with one two-fold dilution above the resistant threshold and several below the susceptible threshold to provide a range for evaluating the results. For example, if interpretive standards are: less than or equal to 1, susceptible (S); 2, intermediate (I); greater than or equal to 4, resistant (R), then the reference plate should include serial two-fold dilutions between 0.25  µg/mL and 8 µg/mL. Including one concentration above the resistant threshold provides data for essential agreement (EA) evaluations. We believe including concentrations more than two dilutions above the resistant concentration provides little evaluable data. Including dilutions below the susceptible category provides more results that are on-scale, and therefore, available for inclusion in the calculations of the EA of evaluable results. The table format samples, Table 5 and 5A, show our recommendations for determining graphically whether a result is evaluable or not.

The reference method performed at the clinical sites may produce errors when testing clinical organisms. Manufacturers may avoid this problem with the challenge and reproducibility assessments by comparing the new device results to a pre-determined expected value, instead of to the results of the reference method performed at the sites. If expected values are not used for the challenge organisms and there is a concern about the variability of the reference method, FDA suggests performing the initial reference test in triplicate for all clinical and challenge organisms. This will help reduce any potential bias. Special care should be taken in the preparation of all reference plates, since the reference result will be used in the final analysis.

B. New Device

Both the new device and the reference panel should include a sufficient number of serial two-fold dilutions around the susceptible and resistant thresholds. For a quantitative minimal inhibitory concentration (MIC) device, the concentrations tested should include at least five two-fold dilutions that surround the susceptible and resistant thresholds of the antimicrobial agent as described above.

C. Organism Selection

You should select organisms for the comparative study that represent the clinical indications of the antimicrobial agent and are within its spectrum of activity according to the Microbiology and Indication and Usage Sections of the FDA approved pharmaceutical antimicrobial agent package insert, and the most recent NCCLS M100 (Ref. 3) Informational Supplement. (See NCCLS M100 Table 1 "Suggested Groupings of U.S. FDA-Approved Antimicrobial Agents That Should be Considered for Routine Testing and Reporting of Non-fastidious Organisms by Clinical Microbiology Laboratories" and Table 1A for fastidious organism recommendations). You should include organisms for which clinical efficacy and in vitro activity have been demonstrated.

A 50% susceptible, 50% resistant distribution within species would be ideal, but such a distribution may be rare when sequential clinical isolates are tested. You should avoid using the same organism from multiple sources and repeat isolates obtained less than three days apart from the same patient. Organisms with known mechanisms of clinically significant resistance should be included in the comparison study as either fresh or selected stock and challenge organisms.

The following example may help to clarify the types of organisms that we recommend you test. If the antimicrobial agent has been shown to be active against Enterobacter cloacae, Klebsiella oxytoca, and Citrobacter spp., both in vitro and in clinical infections, then all Enterobacteriaceae routinely isolated would be relevant for testing.

We recommend you avoid testing organisms that are not included in the Microbiology and Indication and Usage Sections of the FDA approved pharmaceutical antimicrobial agent package insert, because we do not believe such testing provides useful information. If, for example, Pseudomonas spp. are not indicated, they should not be selected for testing.

There are situations where the spectrum of activity (i.e., resistance) of the antimicrobial agent for certain organisms has not been demonstrated in bacteriological or clinical studies. In this instance, the FDA approved pharmaceutical antimicrobial agent package insert has only a susceptible interpretation. Since only organisms in the susceptible category are available, your labeling should recommend that results other than susceptible be referred to a reference laboratory for further analysis. In the event that resistant strains become available, they can be evaluated later, after which you may need to submit a new 510(k).

FDA discourages testing rare organisms for which antimicrobial agents are approved for use, since sufficient data for the rare organism is usually difficult to acquire in a clinical setting. Refer to Tables 1 and 1A in the NCCLS Approved Standard (Ref. 3), for recommended organisms to include or exclude in an evaluation.

1. Fresh clinical organisms

You should include organisms isolated from routine cultures processed in the clinical laboratory study site in the 7 days preceding testing. You should also include all isolates in the appropriate testing group as indicated for testing with the antimicrobial agent in the test device. You should perform the reference method in parallel with the new device.

2. Clinical stock organisms

Generally, each site has its own collection of infrequently isolated or less common organisms. These are organisms saved because of their unique growth or resistance patterns. This selection may be used to enhance the clinical isolates, but should not comprise more than 50% of any group of organisms or the total number tested. You should include these in the study as necessary to incorporate a wider variety of genus and species and also to augment the number of resistant organisms tested.

3. Challenge organisms

You should select challenge organisms from the organisms listed in the Microbiology and Indication and Usage Sections of the FDA approved pharmaceutical antimicrobial agent package insert. The challenge organisms should fall within the spectrum of activity of the antimicrobial agent. The selection of these isolates should favor resistant strains and include organisms for which the antimicrobial agent’s MIC is on-scale. If interpretive criteria are less than or equal to 4 (S), 8 (I), > 8 (R), then we believe organisms with known results in all dilutions between 0.25 and 32 µg/mL are appropriate selections. These challenge organisms are meant to demonstrate whether a device can reliably detect intermediate and resistant organisms. These organisms may be available from the Centers for Disease Control and Prevention (CDC) or a reference laboratory that collects and characterizes strains based on their resistance patterns or particular uniqueness. You may add a selection of organisms that were not used in the developmental stages of the antimicrobial agent algorithm for susceptibility testing, if they are clinically indicated organisms for in vitro testing as stated in the FDA approved pharmaceutical antimicrobial agent package insert.

If the organisms have been characterized phenotypically with repeated NCCLS standard reference method testing, these consensus results can be used as the "expected results." If the "expected result" is not known, you should perform MIC testing using the reference method before using them in the evaluation. You may do so internally or at an outside site. Only the reference method results should be used to determine the expected results.

If the challenge organisms have known expected values, reproducibly obtained using the NCCLS standard reference method, the clinical site need only perform testing with the new device. This will reduce the burden at the clinical site. You should code the set with the expected results, mask it and send it to one site for performance testing on the new device. Alternatively, you could conduct performance testing of the challenge organisms on the new device and the reference method at the same time.

D. Quality Control

We recommend that you conduct the following quality control testing for both the reference method and the new device as well as with any procedural options given in the labeling of the new device:

E. Reproducibility

You should test a minimum of 25 selected organisms with known on-scale results for which the antimicrobial agent is indicated. These may be challenge organisms or other organisms with known results. We recommend that you code organisms and send the organisms to three sites for testing, one time at each site on the new device only. Since this study design will not produce variability within sites, you should provide internal summary data to demonstrate this.

An alternative reproducibility assessment may be performed using 10 selected organisms with known results on-scale. You should not use the quality control isolates if they are not on-scale. These 10 organisms should be tested at each site on three separate days in triplicate with a different inoculum prepared for each test (27 results per isolate). Using this study design, you should calculate reproducibility for within-site (intra-site), for each site, and between-sites (inter-site). See also the table format samples for presenting your reproducibility results, Tables 6A and 6B.

Personnel at each site in the study should perform the same reproducibility study for all inoculum preparation methods and/or reading options recommended in the package labeling. See also Table 1 which outlines in tabular form, FDA’s recommendations for reproducibility testing.

X. Data Presentation

We recommend that you provide summary data for comparative performance, reproducibility, and quality control.5 We have provided several table format samples at the end of this guidance as examples of how to present your results.

A. Comparative Performance Data

Tables 5 and 5A are table format samples that summarize performance using:

The formulas for calculating the percent agreement and the discrepancies listed above are included in Table 5A and in the glossary section.

1. Clinical – fresh and stock

A list of organisms tested should be presented in chart format for each site, identifying the numbers that are stock and fresh for each genus or species. A line listing for all organisms with MIC and/or category result discrepancies between the reference method and the new device should be presented to include genus or species, site, reference method result, test result, type of error, method of inoculation and reading, if applicable. You should submit summary data for all organisms and all study sites combined using Table 5A. This table can also be used to summarize all organisms by site. You should also provide summary data by organism using Table 5.

2. Challenge organisms

You should present results from challenge organisms with the comparison to the expected value or reference result performed at the time of testing. The table format sample in Table 5A is suitable for this purpose. The formulas for calculating the percent agreement and discrepancies are included in Table 5A and in the glossary section.

You should also present all methods of inoculation or reading of results separately. You may wish to use the table format sample in Table 5A for this as well.

We recommend that you also present a line listing of all discrepancies, including the name of the organism, site, reference method result, test result, type of error, and method of inoculation and reading, where applicable.

3. Challenge plus clinical organisms

We recommend that you also present summary data for the challenge data combined with the clinical data. You may wish to use the table format sample Table 5A for this purpose. If there appears to be trending for a particular group of organisms (e.g., Staphylococcus spp., Pseudomonas spp.) such as that observed in the clinical data presented in Table 5 of this guidance, you may present these groups separately. See also Table 3 for an additional table format sample you may wish to use for presenting the EA, CA, discrepancies, and evaluable results in a concise manner for both challenge and clinical data. Table 3 will not demonstrate trending like Table 5, but it will provide an overview for ease in selecting the appropriate organisms for inclusion in the final analysis.

B. Quality Control

Table 4 gives an example of how you should present quality control strain results. We recommend a minimum of 20 test results per site for each method of inoculation and/or reading included in the package insert. You should present both initial and repeat quality control results with an explanation of the action taken for all out-of-range test results.

C. Reproducibility

You should present reproducibility data for all procedural options. If you used the 25-organism study design, you may wish to use the table format sample in Table 6 for this purpose. You should also provide a summary of the internal studies demonstrating the variability across repetitions of the same organism. If you used the 10-organism study design, you may wish to use the table format sample in Table 6A and 6B for this purpose.

With multiple procedural options, the presentation of data is the same for each option.

________________
5 In order to meet the requirements of section 21 CFR 807.87(i), you must include in your 510(k), a financial certification or disclosure statement or both, as required by 21 CFR Part 54. Please refer to Guidance for Industry: Financial Diclosure by Clinical Investigators, issued 03/20/2001, http://www.fda.gov/oc/guidance/financialdis.html.

XI. Evaluating the Results of your Study

The following are recommendations for evaluating the results of your study. The quality control and reproducibility results should also be considered when assessing comparative performance of the reference method and the new device.

A. Fresh, Stock, and Challenge Organisms

Using the table format sample given in Table 5 will help you and FDA visualize discrepancies and trending by organism. However you choose to present the results, you should include only those organisms that would be routinely tested for the antimicrobial agent. For example, you should not include Pseudomonas spp. results for antimicrobial agents that have indications only in the Enterobacteriaceae group, or for Enterococcus, if the indications only include Staphylococcus spp. An additional purpose for the recommendation to use these tables is to identify the on-scale or evaluable test results based on the interpretative criteria of the antimicrobial agent and the concentrations tested on both the reference method and new device.

You should pay particular attention to the organisms with clinical utility and within the spectrum of activity of the antimicrobial agent as shown in the Microbiology and Indication and Usage Sections of the FDA approved pharmaceutical antimicrobial agent package insert. If the EA and CA that you obtain for the organisms listed in the FDA approved antimicrobial agent labeling are below 90%, we recommend that you add a limitation statement to your labeling and consider conducting a future study to support acceptable performance.6 This kind of limitation statement is not necessary for organisms (genus or species) for which the antimicrobial agent has no clinical utility or is inactive and has not been approved for use by FDA (e.g., cefdinir with Enterococcus and Pseudomonas sp.). If these are tested on a panel of antimicrobial agents, NCCLS recommends not reporting results.

You should also evaluate the overall performance of the AST device in the clinical studies for the ability to detect resistance. The use of challenge and stock organisms may be of particular importance in selecting organisms with resistance.

Tables 8 and 9 show the discrepancy rate and minimum acceptable EA rate we recommend. Discrepancy rates will vary depending on the proximity of the MIC of the organisms tested to the interpretative categories. FDA considers the following to be acceptable performance for the clinical data for AST devices for all organisms appropriate for testing:

B. Quality Control

Test results on the new device for the recommended quality control isolates should be within the expected range 95% of the time. In rare events, the expected result with the new device may not agree with the NCCLS recommended ranges for an antimicrobial agent. In this case, you should submit additional data following NCCLS recommendations in M23 "Development of In vitro Susceptibility Testing Criteria and Quality Control Parameters." (Ref. 5). These data should demonstrate the reproducibility of the newly submitted range, plus supportive data showing that all parameters of the test method are in control. These data should include all quality control parameters (e.g., inoculum density check). You should include a statement in the product insert that alerts the user to your unique quality control range. Quality control results that are frequently out of the recommended range will also require a closer scrutiny of the other data to determine if there is a similar trend that might affect clinical results.

If one procedural option provides quality control results that are not accurate for any particular antibiotic while another procedural option produces accurate results (e.g., inoculum preparation, automated reading), you should include a limitation in the labeling stating that results should not be reported for that antibiotic when this particular procedural option is used.

C. Reproducibility

It is difficult for the FDA to determine substantial equivalence for a device if the results of the overall reproducibility study from all test sites for any antimicrobial agent show < 95% (+/- 1 dilution) agreement as compared to the mode. If there is a trending-bias or reproducibility problem with a different procedural option (e.g., inoculum preparation, automated reading), you should include a limitation statement similar to that in section XII,. Labeling, stating that users should not report the results. This type of limitation may apply if some procedural options (method of inoculum, reading method, etc.) were considered unacceptable while another was acceptable.

Observations of trending by a particular organism group or by a procedural option should be investigated further in the other study data to assess the impact on interpretations of patient results.

________________
6 We believe revising this limitation in a legally marketed AST device significantly affects the safety and effectiveness of the device. Therefore, you will need to submit the results of this future study to FDA as a new 510(k) to support modifying this limitation according to section 21 CFR 807.87(g).

XII. Labeling

The premarket notification should include labeling in sufficient detail to satisfy the requirements of 21 CFR 807.87(e). The following suggestions are aimed at assisting you in preparing labeling that satisfies the requirements of 21 CFR 807.87(e).7

The product insert should be flexible to accommodate additional antimicrobial agents. Charts should be used when possible to facilitate additions of future antimicrobial agents, limitations and performance characteristics.

A. Intended Use Statement

The Intended Use statement should indicate:

A typical example of an intended use statement is:

"ABC's system is intended for the in vitro qualitative or quantitative determination of antimicrobial susceptibility of rapidly growing aerobic non-fastidious Gram positive and Gram negative organisms on the ABC Instrument."

B. Reagents

You should list antimicrobial agents along with concentration ranges and abbreviations. You may include these in the reagent section of the labeling or on each package container if different for different devices. To prevent confusion between different drugs with similar generic or trade names, you should use abbreviations recommended by the pharmaceutical manufacturer.

C. Reporting of Results

You should provide the interpretive criteria users should use for each antimicrobial agent on the MIC or breakpoint device. You should present the NCCLS/FDA interpretive standards that you used in the evaluation. AST systems may be able to provide results for organisms that may not be appropriate for all of the antimicrobial agents provided on a test panel or system. Therefore we recommend you explain the clinical utility of your interpretive criteria in your labeling. For example:

"There are antimicrobial agents included in this [panel, device, or section] that have not been proven to be effective for treating infections for all organisms tested. Refer to the individual FDA approved pharmaceutical antimicrobial agent package insert for interpreting and reporting results of antimicrobial agents that have shown to be active against organism groups both in vitro and in clinical infections."

Alternatively, you could refer to the most recent NCCLS M100 Informational Supplement, Tables 1 and 1A. If manual readings are an option, provide a chart of what thresholds should be used for SIR (Susceptible, Intermediate, Resistant) interpretations or refer to the NCCLS M100 Tables 1 and 1A for appropriate organism reporting with interpretations in Table 2 of the same NCCLS document.

Results should not be reported in instances where performance has not been established either because there are no interpretive criteria for a particular organism group or insufficient numbers of organism groups have been tested. Where feasible, FDA suggests that suppression of results be software driven. Interpretations (i.e., SIR) should not be reported for these groups of organisms. If you report MIC results, they should carry a disclaimer that device performance or antimicrobial agent clinical effectiveness have not been established. MIC results for this type of organism may be useful for antibiogram patterns, but the practice of reporting results should be discouraged when the antimicrobial agent has not been proven to be effective for treating infections caused by these organisms and the performance on your device has not been established.

D. Performance Characteristics

You should describe the study design, stating the reference method used, number of sites, etc. You should list the percent EA and/or CA in table format with the NCCLS standard reference method for each antimicrobial agent from comparative performance evaluations. You should also include results of reproducibility studies in either a table format or a summary paragraph describing the type of study and a statement that all reproducibility results were acceptable at greater than or equal to 95%.

E. Quality Control

You should list all recommended quality control strains whether NCCLS or other and the expected results when tested with each antimicrobial agent.

F. Limitations

You should list all applicable limitations. If the device has software-generated interpretations, these limitations should be incorporated into the software. The following are examples of some limitation statements that may apply to your device:

______________
7 Although final labeling is not required for 510(k) clearance, final labeling must comply with the requirements of 21 CFR 809.10 before an in vitro diagnostic device is introduced into interstate commerce. In addition, final labeling for prescription medical devices must comply with 21 CFR 801.109. Labeling recommendations in this guidance are consistent with the requirements of part 801 and section 809.10.

XIII. Removing Certain Labeling Limitations from Legally Marketed AST Devices

The FDA will request appropriate data and information in a new 510(k) to support the removal of any limitation included in the labeling that was placed there during the original clinical studies. These types of data are described below and detailed in Table 2. You should refer to the submission in which the limitation statement was made. If you made changes to the device to alter the performance, your additional studies should include all organisms previously tested if these changes significantly affect safety and effectiveness.

A. Performance

If a limitation in the labeling is a result of performance characteristics that are based on EA and/or CA and you wish to modify the device in order to delete the limitation, you should perform a comparative clinical laboratory study. This study should follow the design for the comparative study described in this guidance. The organism mix should be concentrated around those groups that provided the original EA or CA results. You should include all groups that might be affected by modifications to the device. You may wish to report the results using the table format samples provided in this guidance.

B. Insufficient Resistant Strains

If a limitation in the labeling is a result of not testing sufficient resistant strains, you should perform a comparative study to demonstrate that the device can detect resistance in organisms that are included in the Microbiology and Indication and Usage Sections of the FDA approved pharmaceutical antimicrobial agent package insert. This testing should utilize reference and test devices similar to those from the original comparative study. A special challenge set containing the resistant isolates and some susceptible organisms may be substituted for fresh isolates. You may wish to report the results using the table format samples provided in this guidance.

C. Reproducibility

If the reproducibility was <95% for a particular procedural option, you should perform a study to verify that this method is now reproducible. This study should include the problematic organism(s) or procedural options (alternate methods of inoculation, alternate reading procedures, etc.) which originally did not demonstrate reproducible results. You should conduct either the 25-organism study, or the 10-organism study at three test sites as described in this guidance. The strains should include organisms for which the antimicrobial agent is intended for testing with known results near the interpretive criteria range. Include organisms you determined to be problematic in the original reproducibility study. You may wish to report the results using the table format samples in Tables 6, 6A, and 6B. You should also include the testing of all quality control organisms. Notable bias or poor reproducibility of an alternate method of inoculation or reading may indicate additional concerns with this particular procedure. Additional challenge data may be needed to resolve these concerns. If you determine that the inoculum was the problem, you should perform an evaluation of colony count data.

D. Quality Control

If alternate methods of inoculation or reading produced quality control values that did not match NCCLS acceptable ranges, you should test a minimum of 20 replicates per site with each quality control organism on the test device to verify that the quality control values are now within the acceptable NCCLS quality control range. Prepare each quality control organism from a different inoculum suspension. You may wish to report the results using the table format samples in Table 4. You should perform colony counts if you do not use a standardized inoculation method (e.g., photometric device). Colony counts should be done once on each day of testing using the NCCLS recommendations for sampling from the inoculated test device. If the device package insert recommends additional methods of inoculation and/or reading, you should test all options. If you elect to propose an alternative range, you should follow an NCCLS M23 (Ref. 5) study design. You should explain any affect on clinical isolate results.

XIV. QSR Considerations

Part of the QSR (Quality System Regulation, 21 CFR Part 820) is to ensure that the finished product will be safe and effective and perform as intended. There is a concern to the public health when AST devices can not reproducibly generate the same results within +/- one well. If the non-reproducible result is around the interpretative criteria, a vmj could occur. The patient report would recommend for treatment an antibiotic to which the organism is actually resistant, which could lead to treatment failure, particularly for serious infections or altered host conditions. Another possibility is a major error, which can lead to utilization of broad-spectrum agents and needlessly accentuate the pressure for selection of resistant flora. This may also result in further risk to patients because the selection of treatment antibiotic may now be an antibiotic that could be more toxic to the patient when a less toxic antibiotic is available.

You should consider the reproducibility and stability of components in the design of the device and in the development of release criteria. All aspects of the final product will have an effect on the performance, but the lot-to-lot reproducibility and stability studies of the antimicrobial agent should be for performance of the antimicrobial agent only. You should validate the other aspects by other means or at different times. Testing should be performed internally with design components similar to the clinical trial protocol. You should keep the data on file and available upon request.

Manufacturers are also expected to use surveillance data and customer complaints as part of design controls (21 CFR 820.100) Corrective and preventive action). The NCCLS includes surveillance data in performing continuous assessment of older antimicrobial agents, particularly when new mechanisms of resistance emerge. The manufacturer is responsible for keeping abreast of all information to better reevaluate the product to determine if the change in resistance patterns has affected performance and accuracy of the test.

Although not part of the class II special controls, FDA recommends that you consider the following in your approach to complying with QSRs. The method you use to validate reproducibility and stability should be able to detect a change in potency of at least 50% for each antibiotic. For example, if the organisms you select for such testing have stable on-scale MICs, then a sufficient number of replicates of these organisms could detect shifts in the mode of the test organism. Although FDA acknowledges that the methodology has a +/- one well variability, each antimicrobial agent should be evaluated by you for even slight trending to ensure the product will continue to be safe and effective and perform as intended.

A. Lot to Lot Reproducibility

Your study design should demonstrate that different lots of prepared antimicrobial agents in the final format (minimum 3 lots) will perform with the same accuracy. If all other device components have been previously evaluated, you need only include one lot of these components since this study design is to monitor the antimicrobial agent.

B. Stability

Your study design should verify the shelf life of the antimicrobial agents in their final format for all conditions that are recommended by your labeling. The temperatures at which the product is stored should include the extremes of the range recommended for storage. Include observations of slight trending in one direction over time as part of the evaluation. Your study design should verify the shelf life of the antimicrobial agents in their final format for all conditions that are recommended by your labeling. The temperatures at which the product is stored should include the extremes of the range recommended for storage. Include observations of slight trending in one direction over time as part of the evaluation.

XV. Glossary

Agar Dilution Susceptibility Test An antimicrobial susceptibility test method using concentrations of an antimicrobial agent incorporated into agar growth medium plates.

Agreement - Category (CA)

Agreement of interpretive results (SIR) between a new device under evaluation and an NCCLS standard reference method. FDA interpretive criteria should be used but the NCCLS criteria, if different, may also be applied if based on more current recommendations for detecting organism resistance when resistant mechanisms were not recognized (or did not exist) during the FDA drug evaluation.

Agreement, Essential (EA)

Agreement within plus or minus, one two-fold dilution of the new device under evaluation with the reference method MIC determination. Both the new device and the reference method should test a range of two-fold dilutions that include at least one dilution above and below the interpretive thresholds.

Bias

Measure of whether the new device produces the correct result.

Breakpoint System

Systems similar in design to MIC systems, but with four or fewer concentrations of each antimicrobial agent. These concentrations are the interpretive thresholds (based on the FDA or NCCLS interpretive categorical MIC values for each antimicrobial agent) that provide a qualitative (category) result (SIR). FDA considers these devices qualitative.

Broth Dilution Susceptibility Test

An antimicrobial susceptibility test method using concentrations of an antimicrobial agent in broth growth medium, in either tubes (macrodilution) or wells (microdilution).

Discrepancy

A disagreement between the new device result and the reference method. Either the new device MIC is greater than plus or minus one two-fold serial dilution and/or the interpretive category is different.

Discrepancy - major (maj)

The NCCLS reference category result is S and the new device result is R. To calculate the major discrepancy rate, use the following formula:

Discrepancy major equation


Discrepancy – minor (min)

The NCCLS reference category result is R or S and the new device result is I; or the reference result is I and the new device result is R or S. To calculate the minor discrepancy rate, use the following formula:

Discrepancy minor equation


Discrepancy – very major (vmj)

The NCCLS reference category result is R and the new device result is S. To calculate the very major rate, use the following formula:

Discrepancy very major equation


Evaluable Result

When the reference method result is on-scale and the new device result is also on-scale. FDA believes that if the reference result is on-scale and the new device result is not on-scale, comparative data may not be evaluable. FDA does not consider evaluable any reference result that falls in the less than or greater than category. However, such results may be part of the EA and/or CA assessments. See Table 5 and 5A for examples.

Evaluable (i.e., on-scale) results are those that fall within the test range of the reference method and could also be on-scale with the new device if within the plus/minus one well variability.


Fastidious Organism

Those that require very specialized nutrients and environmental conditions to thrive and remain viable. For the purposes of this document, a fastidious organism is one that will not grow well in (or on) unsupplemented Mueller-Hinton medium within 24 hours.

Genotypic Resistance

The presence of resistance-expressing genes. The presence of resistance-expressing genes can often infer resistance. Absence of these resistance determinants cannot generally exclude resistance by other mechanisms.

In vitro Diagnostic
(IVD)

In vitro
diagnostic products (reagents, instruments, and systems) that are medical devices under the Federal Food, Drug, and Cosmetic Act. The generic product class is intended for use in clinical laboratories for determining in vitro susceptibility or resistance of bacterial pathogens to therapeutic agents. (see 21 CFR Section 866.1640).

Inoculum Density Check

Plate counts performed to ensure that the numbers of organism inoculated into the test system are within prescribed ranges. See NCCLS M7 Approved Standard (Ref. 1).

Minimal Inhibitory Concentration (MIC)

The lowest concentration of an antimicrobial agent that prevents visible growth of a microorganism in an agar or broth dilution susceptibility test.

Minimal Inhibitory Concentration (MIC) Systems

Broth dilution, agar dilution, or other methods or systems that have at least five concentrations of generally two-fold dilutions of antimicrobial agents. These may be in broth, plate, gradient diffusion, or other formats. The antimicrobial concentrations may be frozen, lyophilized, or dehydrated. They should include a minimum of two dilutions below the susceptible threshold in order to assess developing resistance, and to trend and track patterns of resistance. These devices provide quantitative MIC results. They can be manual, semi-automated, or fully automated.

Mode

The most frequently occurring MIC result when one organism is repeatedly tested.

On-scale Result

An MIC result from testing a series of dilutions when there is growth in at least one, but not all concentrations tested.

Organism - Challenge

Selected organisms with expected MICs that are near the SIR thresholds or at least on-scale. Testing these challenge organisms enriches the numbers of organisms in the evaluation with evaluable results.

Organism – Clinical stock

An organism isolated from a clinical specimen at a clinical laboratory site, has been retained/stored for more than 7 days, and is used in the comparative study between a new device and the NCCLS standard reference method. These are usually retained because they have: known mechanisms of resistance, have an unusual susceptibility pattern to antimicrobial agents in the same class as the antimicrobial agent under evaluation, or are the genus and/or species for which the antimicrobial is indicated but are not commonly isolated, and would likely not be included with fresh organisms used in the evaluation.

Organism - Fresh clinical

Organisms isolated from clinical specimens at a clinical laboratory site during 7 days prior to testing in the comparative evaluation between the new device and the reference method. These organisms should not be frozen or repeatedly subcultured.

Phenotypic Resistance

Observable or measurable in vitro growth in the presence of a known antimicrobial concentration.

Predicate

A device that was legally marketed prior to May 28, 1976 (preamendments device), or a device which has been reclassified from Class III to Class II or I, or a device which has been found to be substantially equivalent to such a device through premarket notification.

Procedural Options

Optional methods in the instructions for use (Procedure Section) in the package insert for the new device. Examples of such procedural options are: alternate organism inoculation preparation methods such as direct colony suspension without turbidimetric qualification, visual reading when the system is primarily instrument-read and automated readings.

Purity Check

A quality control procedure to ensure that the growth endpoint for an MIC or breakpoint result is not caused by more than one organism. See NCCLS M7 Approved Standard (Ref. 1).

Qualitative Susceptibility Result

A category result (S, I or R) obtained with a device containing four or fewer concentrations of an antimicrobial agent.

Quantitative Susceptibility Result

An MIC result obtained with a device containing five or more concentrations of an antimicrobial agent. In addition to reporting a category result of susceptible (S), intermediate (I), or resistant (R), the actual MIC can also be reported.

Reference Method

Standard broth dilution (macrodilution or microdilution) or agar dilution as described in NCCLS M7 Approved Standard (Ref. 1).

Reproducibility

Measure of whether the new device produces the same result across different testing conditions.

Resistant Threshold

Highest in vitro concentration at which most organisms are no longer considered susceptible. Organisms with an MIC at this concentration, or higher are reported as resistant.

Shortened Incubation

Determinations of growth in less than 16 hours.

SIR

Susceptible, Intermediate, Resistant.

Susceptible Threshold

Lowest in vitro concentration at which most organisms are still considered susceptible. Organisms that do not grow at this concentration or at lower concentrations are reported as susceptible.

Trending

An upward or downward change associated with increased resistance (decreased susceptibility) or increased susceptibility (decreased resistance). This type of change may not necessarily be seen with qualitative susceptibility testing. Trending is applied for certain organisms or certain antimicrobials to detect emerging resistance or may be used to compare results between different susceptibility testing methods to assess bias that would not be evident using EA or CA, unless larger numbers of organisms were evaluated.

XVI. References

1. NCCLS. M7 - Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically; Approved Standard. (most recently approved supplement). NCCLS; Wayne, Pennsylvania.

2.

NCCLS. M11 - Methods for Antimicrobial Susceptibility Testing of Anaerobic Bacteria; Approved Standard. (most recently approved supplement). NCCLS; Wayne, Pennsylvania.

3.

NCCLS. M100 - Performance Standards for Antimicrobial Susceptibility Testing. (most recent informational supplement). NCCLS; Wayne, Pennsylvania.

4.

Ferraro MJ, Jorgensen JH. Susceptibility Testing in Instrumentation and Computerized Expert Systems for Data Analysis and Interpretation. In: Murray PR, Baron EJ, Pfaller MA, et al, eds. Manual of Clinical Microbiology, 7th Edition, Washington DC: American Society of Microbiology; 1999 1593 – 1600.

5.

NCCLS. M23 - Development of In vitro Susceptibility Testing Criteria and Quality Control Parameters; Approved Guideline. (most recently approved supplement). NCCLS; Wayne, Pennsylvania.

TABLE 1: Recommendations for Antimicrobial Susceptibility Devices  a

 

MIC/BP
Formats

Fastidiousb

Additional methods of
Inoculation/Readingc

Number of Sites (including 1 in-house)

3

3

3

Organisms Freshd Clinical/stocke

100/site

75/site

0

CDC Challengef

75/one site

50/one site

75 or 50/one site

Reproducibilityg

25/site or 10x3x3/site

25/site or 10x3x3/site

25/site or 10x3x3/site

Interpretive Standards

FDA/NCCLS

FDA/NCCLS

FDA/NCCLS

Stability (3 lots)

Real time (on file)

Real time (on file)

Real time (on file)

QC Reference and
Test Device Results
NCCLS Strains

20 results/site

20 results/site

20 results/site

(Other Mfg.
Recommended)

Optional

Optional

Optional

On-scale

At least 1

At least 1

At least 1

Inoculum density checkh

QC, reproducibility, fresh

QC, reproducibility, fresh

QC, reproducibility

NCCLS Reference Method

MIC

MIC

MIC

_____________________
a See Tables 5, 8, and 9 for statistical calculation examples and evaluable results.
b For Fastidious organisms such as Streptococcus, Haemophilus, anaerobes, etc. that have an NCCLS approved standard methodology, FDA/NCCLS interpretive criteria and quality control recommendations, refer to NCCLS approved standard M100 (Ref. 3) Table 1A. The routine testing of rare isolates is not recommended.
c Minimal data to establish performance should be presented for each procedural option of the method of inoculation (growth, direct colony suspension etc.), reading of results (visual vs. automated), or combinations of options.
d Fresh clinical - an organism isolated from a clinical specimen which has been on an agar plate for less than 7 days and not frozen.
e Stock organisms - any organism from a clinical specimen which has been isolated greater than 7 days prior to testing. Should not include organisms for which the antimicrobial agent is not intended. Selection should be supplemental based on the listing in the FDA approved pharmaceutical antimicrobial agent package insert and should not comprise more than 50% of the clinical isolates.
f Challenge - CDC or reference laboratory source with known results (preferably on-scale) to be tested on the test system. Organisms that are intended for the testing with the antimicrobial agent as stated in the FDA approved pharmaceutical antimicrobial agent package insert (microbiology section), should be selected for testing on the test device.
g All on-scale results.
h Inoculum density check should be performed daily on the QC isolates, on reproducibility isolates, and 10% of fresh isolates. Alternate approaches may be substituted if the inoculum method uses a spectrophotometric device.

TABLE 2: Recommendations for the Removal of Limitations from Antimicrobial Susceptibility Devices a

Items

Performance

Insufficient Resistant
Strains

Quality Controlb

Reproducibility

Number of Sitesc

3

1

3

3

Organisms

Fresh or
Stock
Clinical

100/site

75d

0

NA

Challenge

75/one site

as needed

0

 
Reproducibilitye

NA

NA

NA

25/site or 10x3x3/site

Quality Control

Daily

Daily

20

Daily

__________________
a For statistically evaluable numbers see Tables 5, 8, and 9
b To be used to demonstrate that the QC ranges are now in the same ranges as the FDA/NCCLS
c One may be in-house
d A minimum of 75 organisms either resistant or clustered near the susceptible threshold.
e On-scale

Note:
1. If changes have been made to the device to alter the overall performance, the testing should include all organisms previously tested.
    Refer to Table 1.
2. Perform testing for all procedural options.

TABLE 3: Presentation of Summary Data for Both Challenge and Clinical Data

Clinical Data

Organism Group

Total
Tested

# EA

%EA

Total
Evaluable

# EA of
Evaluable

%EA of Evaluable

# CA

%CA

# R

# vmj

# maj # min
K. pneumoniae 79 74 93.7 64 62 92.2 70 88.6 20 0 1 8
P. aeruginosa 96 90 93.8 91 85 93.4 88 91.7 40 0 1 7
C. freundii 26 21 80.8 18 14 72.2 23 88.5 10 0 0 3
E. aerogenes 22 21 95.5 21 21 95.2 21 95.5 2 0 0 1
E. cloacae 57 53 93 50 48 92 51 89.5 13 1 1 4
M. morganii 15 14 93.3 12 12 91.7 15 100 7 0 0 0
P. mirabilis 34 26 76.5 34 29 76.5 34 100 33 0 0 0
E. coli 92 85 92.4 62 59 88.7 87 94.6 43 2 1 2
S. marcescens 50 49 98 48 48 100 47 32 32 0 0 3
Acinetobacter spp 41 40 97.6 29 28 96.6 34 14 14 0 0 7
TOTAL 512 473 92.4 429 406 94.6 470 91.8 214 3 4 35

Challenge

K. pneumoniae 10 10 100 10 10 100 10 100 6 0 0 2
P. aeruginosa 20 18 90 17 15 88.2 19 95 15 1 0 0
C. freundii 10 10 100 9 9 100 9 90 2 0 1 0
E. aerogenes 5 4 80 4 3 75 5 100 0 0 1 1
E. cloacae 5 5 100 2 2 100 5 100 4 0 0 1
Aeromonas sp. 5 4 80 4 4 100 5 100 0 0 0 0
P. mirabilis 10 9 90 8 6 75 9 90 4 0 0 2
S. marcescens 10 10 80 9 9 100 9 90 2 0 0 1
Acinetobacter spp 10 8 80 5 5 100 8 80 2 1 0 2
TOTAL 85 78 91.8 68 63 92.6 79 92.9 35 2 2 9

Clinical and Challenge Combined

All organisms 597 551 92.3 497 469 94.4 549 92.0 249 5 6 44

TABLE 4: Example of Reporting Format for Quality Control Data

Antimicrobial agent: ___________________________

QC Organism Expected Result

Reference Result Frequency

New Device Frequency

E. coli ATCC 25922 0.25 - 1.0 µg/mL  

Site 1

Site 2

Site 3

Site 1

Site 2

Site 3

<.25      

1

1

 
.25

14

18

14

4

14

14

.5

6

2

6

4

5

6

1.0      

11

   
>1.0            
E. cloacae Ref 1611 2 - 8 µg/mL <2            
2      

12

14

4

4

14

15

12

2

6

5

8

6

5

8

8

 

11

>8            
Pseudomonas aeruginosa ATCC 27853 0.5 - 2 µg/mL <.25    

2

   

2

.5

10

10

18

 

2

 
1

5

8

2

20

18

18

2

5

2

     

4

4            
Enterococcus faecalis ATCC 29212 4 - 16 µg/mL 2            
4

18

2

18

20

12

6

8

2

18

2

 

8

14

16            

TABLE 5: Sample Table Format for Device Performance

Antimicrobic agent: Oxacillin Organism: Staphylococcus aureus

Test Results

Reference Results

Evaluation

 

lesser than or equal to 0.25a 0.5 1 2 S 4 R 8 greater than or equal to 16

Overall EAb

199/208

95.7%
lesser than or equal to 0.25 6a a 1       a
0.5 102 100 21 2     a EA based on evaluable resultsa b

152/161

94.4%
1 a 10 8 1     a
2 S a 6 11 1     a CA based on interpretationc   100%
4 R a           a
greater than or equal to 8 a       a a 31a
 
Evaluable Results   116 41 4      

Antimicrobic agent: Oxacillin

Organism: Staphylococcus epidermidis

Test Results

Reference Results

Evaluation
  lesser than or equal to 0.25 S 0.5 R 1 2 4 8 greater than or equal to 16 Overall EAb 108/116 93.1%
 lesser than or equal to0.25 S 29a a         a
0.5 R a           a EA based on evaluable results a b 3/11 27.3%
1 a           a
2 a   1 1   1 a CA based on interpretation c

 

100%
4 a       1   a
greater than or equal to 8 a     7 12a 12a 52a
 
Evaluable Results     1 8 1 1  

Antimicrobic agent:
Oxacillin

Organism:
other CNS

Test Results

Reference Results

Evaluation
  lesser than or equal to 0.25 S 0.5 R 1 2 4 8 greater than or equal to 16 Overall EA b 67/70 95.7%
lesser than or equal to 0.25 S 25a 1a vmj         1a
0.5 R a 8 3       a EA based on evaluable results a b 11/11 100%
1 a           1a
2 a           a CA based on interpretation c 67/70 95.7%
4 a           a
8 1a (maj)       2a 3a 26a
 
Evaluable Results   8 3        

see footnotes on Table 5A

Present all fresh and stock results for organisms in a genus/species in a separate diagonal. For example all Staphylococcus aureus results from all sites would appear in the first diagonal.

TABLE 5A: Sample Table Format for Device Performance, continued

Antimicrobic agent: Ciprofloxacin Organism: All

Test Results

Reference Results

Evaluation
  lesser than or equal to 0.12 0.25 0.5 1 S 2 I 4 R lesser than or equal to 8

Overall EA b

398/407 97.8%
lesser than or equal to 0.12 259a a         a
0.25 a 4 2       a EA based on evaluable results a b 85/88 96.6%
0.5 6a 4 2       a
1 S a   10 10 5 1 a CA based on interpretation c 371/407 91.2%
2 I a   2 9 10 11 a
4 R a       8 10 a CA Minor d 35/407 8.6%
greater than or equal to 8 a         1a 53a Major e 0 0%
Evaluable Results   8 16 19 23 22   Very major f 1/76 1.3%
________________________
a Results are not considered evaluable if they fall in this box
(b) Essential Agreement (EA) Equation
(c) Category Agreement (CA) Equation
(d) Minor Discrepancy (min) Equation
(e) Major Discrepancy (maj) Equation
(f) Very Major Discrepancy (vmj) Equation

To be used for:

TABLE 6: Presentation of Reproducibility Results a by Organism

 

Difference in the number of wells between test result and test mode

 

 

Off-scale

-2

-1

0

+1

+2

Off-scale

Site 1 result

Site 2 result

Site 3 result

Test Modeb

                       
P. aeruginosa      

3

     

0.5

0.5

0.5

0.5

P. aeruginosa      

2

1

   

4

2

2

2

P. aeruginosa    

1

2

     

16

32

32

32

P. aeruginosa      

2

1

   

8

16

8

8

E. coli      

3

     

2

2

2

2

E. coli    

1

2

     

8

16

16

16

E. coli      

2

1

   

4

2

2

2

E. coli      

3

     

1

1

1

1

E. coli    

1

2

     

2

4

4

4

E. coli      

3

     

32

32

32

32

M. morganii    

1

2

     

16

16

8

16

C. diversus      

2

1

   

0.5

0.5

1

0.5

C. freundii      

2

1

   

1

2

1

1

C. freundii      

3

     

16

16

16

16

E. cloacae      

2

1

   

16

32

16

16

E. cloacae      

2

 

1

 

0.5

0.5

2

0.5

E. cloacae      

3

     

4

4

4

4

P. mirabilis      

3

     

8

8

8

8

P. mirabilis      

3

     

2

2

2

2

S. marcescens      

2

1

   

2

1

1

1

S. marcescens      

2

1

   

32

16

16

16

S. marcescens      

2

1

   

4

2

2

2

K. pneumoniae      

2

   

1

32

32

>32

32

K. pneumoniae      

2

1

   

0.25

0.5

0.25

0.25

P. stuartii  

1

 

2

     

16

16

4

16

                       
Total  

1

4

58

10

1

1

       
Between-site reproducibilityc

73/75 = 97.3%d
72/75 = 96%e

______________________
a Results in the table are occurrences of the difference in the number of wells between the test result and the test mode. The study is based on 25 on-scale organisms, tested at 3 sites.
b Most frequent new test result. If there is no mode, the median should be used.
c Total number of results that fall within 1 well (+/-1) of the mode result divided by total number of results. This should be calculated for the best and worst case if some of the values are off-scale while some are on-scale. If all three results are off-scale they should still be included in the calculation as part of the best-worst case calculations. For this study the denominator would always be 75 whether the results are off or on-scale.
d Best case calculation for reproducibility assuming the off-scale result is within one well from the mode.
e Worst case calculation for reproducibility assuming the off-scale result is greater than one well from the mode.

TABLE 6A: Presentation of Reproducibility Results a by Organism and Site

 

Difference in the number of wells between new test result and test mode

 

 

Off-scale

-2

-1

0

+1

+2

Off-scale

Test Modeb

       

Site 1

       
                 
E. coli 1    

6

3

     

0.5

E. coli 2      

9

     

2

E. coli 3    

6

3

     

2

Pseudomonas 4      

3

6

   

16

Pseudomonas 5    

3

5

   

1

32

Klebsiella 6      

8

 

1

 

8

Enterobacter 7      

8

1

   

8

Serratia 8      

4

5

   

4

Serratia 9      

5

4

   

16

Proteus 10    

5

4

     

1

                 
Total    

20

52

16

1

1

 
Within-Site Reproducibilityc

89/90 = 98.9%d
88/90 = 97.8%e

                 
                 
       

Site 2

       
                 
E. coli 1    

3

6

     

0.5

E. coli 2      

9

     

2

etc…                

________________
a Results in the table are occurrences of the difference in the number of wells between the test result and the test mode. The study is based on 10 on-scale organisms, tested in triplicate on 3 separate days at 3 sites.
b Most frequent test result.
c Total number of results that fall within 1 well (+/-1) of the mode result divided by total number of results. This should be calculated for the best and worst case if some of the values are off-scale while some are on-scale. If all results are off-scale they should still be included in the calculation. For this study the denominator would be ninety whether they are all on scale or not.
d Best case calculation for reproducibility assuming the off-scale result is within one well from the mode.
e Worst case calculation for reproducibility assuming the off-scale result is greater than one well from the mode.

TABLE 6B: Presentation of Reproducibility Results a by Organism, Pooled Across Sites

 

Difference in the number of wells between new test result and test mode

 

 

Off-scale

-2

-1

0

+1

+2

Off-scale

Test Modeb

All Sites

                 
E. coli 1    

9

18

     

0.5

E. coli 2      

25

2

   

2

E. coli 3  

3

9

12

3

   

2

Pseudomonas 4    

6

12

9

   

16

Pseudomonas 5    

2

17

   

8

32

Klebsiella 6    

7

14

5

1

 

8

Enterobacter 7    

5

14

8

   

8

Serratia 8    

4

15

8

   

4

Serratia 9    

3

20

4

   

16

Proteus 10    

3

14

10

   

1

                 
Total  

3

48

161

49

1

8

 
Between-site reproducibilityc

266/270 = 98.5%d
258/270 = 95.6%e

___________________
aResults in the table are occurrences of the difference in the number of wells between the test result and the test mode. The study is based on 10 on-scale organisms, tested in triplicate on 3 separate days at 3 sites.
b Most frequent test result.
cTotal number of results that fall within 1 well (+/-1) of the mode result divided by Total number of results. This should be calculated for the best and worst case if some of the values are off-scale while some are on-scale. If all results are off-scale they should be included in the calculation with a denominator of 270 for all calculations.
d Best case calculation for reproducibility assuming the off-scale result is within one well from the mode.
e Worst case calculation for reproducibility assuming the off-scale result is greater than one well from the mode.

TABLE 7: Report Format for Inoculum Density

ORGANISMa NUMBER TESTED SOURCE METHODb MEAN MINUMUM MAXIMUM STD. DEV.
S. aureus
ATCC #
20 QC ATCC Reference 6 X 105 2 x 105 9 x 105  
S. aureus
ATCC #
20 QC ATCC Direct inoculum 5 x 105 2 x 105 8 x 105  
S. aureus
ATCC #
20 QC ATCC Growth inoculum 5 x 105 2 x 105 6 x 105  
MRSA 13 Reproducibility, clinical Direct inoculum 7 x 105 4 x 105 6 x 106  
MRSA 13 Reproducibility, clinical Growth inoculum 6 X 105 2 x 105 8 x 106  
MSSE 3 Reproducibility Direct inoculum 8 x 105 5 x 105 7 x 105  
MSSE 3 Reproducibility Growth inoculum 7 x 105 4 x 105 12 x 105  
MRSE 19 Reproducibility Direct inoculum 6 X 105 2 x 105 8 x 106  
MRSE 19 Reproducibility Growth inoculum 7 x 105 5 x 105 7 x 105  
Enterococcus 4 Clinical Direct inoculum     9 x 105  
Enterococcus 4 Clinical Growth inoculum        
MSSA 15 Clinical Direct inoculum        
MSSA 15 Clinical Growth inoculum        
      Direct inoculum        
      Growth inoculum        

_________________
a Data should be available upon request for "by site" evaluation, by organism, etc.
b Inoculum density should be performed on all methods of inoculation unless a standardized method (photometric device) is used.

TABLE 8: Number of VMJ Discrepancies as a Function of the Number of Resistant Organisms Tested

Number of Resistant Organisms

Acceptable
Number of Discrepancies

Estimated
Ratea

95% Confidence
Intervalb for True VMJ
Rate

48

0

0.00

(0.00, 7.40)

50

0

0.00

(0.00, 7.11)

60

0

0.00

(0.00, 5.96)

70

0

0.00

(0.00, 5.13)

72

1

1.39

(0.04, 7.50)

80

1

1.25

(0.03, 6.77)

90

1

1.11

(0.03, 6.04)

94

2

2.13

(0.26, 7.48)

100

2

2.00

(0.24, 7.04)

110

2

1.82

(0.22, 6.41)

120

3

2.50

(0.52, 7.13)

130

3

2.31

(0.48, 6.60)

140

4

2.86

(0.78, 7.15)

150

4

2.67

(0.73, 6.69)

160

5

3.13

(1.00, 7.20)

170

5

2.94

(0.94, 6.78)

180

6

3.33

(1.21, 7.16)

190

7

3.68

(1.48, 7.48)

200

7

3.50

(1.40, 7.12)

250

8

3.20

(1.38, 6.24)

300

9

3.00

(1.37, 5.64)

400

11

2.75

(1.37, 4.88)

___________________
a Est. Rate = estimated vmj rate = number of vmj discrepancies divided by number of resistant organisms.
b Exact confidence intervals based on the binomial distribution.

TABLE 9: Essential Agreement as Function of the Number of Evaluable Organisms Tested

Number of
Evaluablea Organisms

Acceptable Number of Disagreements

Estimated Essential
Agreement (EA)b

95% Confidence Intervalc for True EA

35

0

100.00 %

(90.00, 100.00)

54

1

98.15

(90.11, 99.95)

55

1

98.18

(90.28, 99.95)

60

1

98.33

(91.06, 99.96)

65

1

98.46

(91.72, 99.96)

70

2

97.14

(90.06, 99.65)

75

2

97.33

(90.70, 99.68)

80

2

97.50

(91.26, 99.70)

85

3

96.47

(90.03, 99.27)

90

3

96.67

(90.57, 99.31)

95

3

96.84

(91.05, 99.34)

100

4

96.00

(90.07, 98.90)

110

4

96.36

(90.95, 99.00)

120

5

95.83

(90.54, 98.63)

130

6

95.38

(90.22, 98.29)

140

6

95.71

(90.91, 98.41)

150

7

95.33

(90.62, 98.10)

160

8

95.00

(90.39, 97.82)

170

9

94.71

(90.19, 97.55)

180

10

94.44

(90.02, 97.30)

190

10

94.74

(90.53, 97.45)

200

11

94.50

(90.37, 97.22)

___________________
a Evaluable (e.g., on-scale) organisms are those that fall within the test range of the reference and have the opportunity for a result on the test method that could also be on-scale. Any reference result that falls in the < or > category is considered not evaluable.
b Estimated Essential Agreement = percent agreement = number of evaluable test results that are equal to or with in one dilution of the expected result divided by number of organisms that are evaluable.
c Exact confidence intervals based on the binomial distribution.

Updated 2/12/2003