FDA
Home Page | CDRH Home Page | Search
| CDRH
A-Z Index | Contact CDRH
|
Medical Device Recalls
|
Date Recall Initiated |
June 3, 2004 |
---|---|
Product: |
Carl Zeiss Ophthalmic System – VISULAS 532s with VISULINK 532/U surgical laser instrument. |
Use: |
This medical device is intended for use in laser treatment of diseases of the eye, particularly in treating retinal detachments or bleeding of the retina. |
Recalling Firm: |
Carl Zeiss Meditec 5160 Hacienda Drive Dublin, California 94568-7562 |
Reason for Recall: |
A mirror is used to direct the laser treatment beam to the patient’s eye. The coating of the reflective mirror of the VISULINK 532/U could tear and loosen from the mirror surface. |
Public Contact: |
Carl Zeiss Meditec Customer Service 800-341-6968 |
FDA District: |
San Francisco |
FDA Comment: |
The faulty mirror may misdirect the laser beam to an unintended target in or on the eye resulting in retinal bleeding and/or burns due to excessive laser energy in the eye. Class I recalls are the most serious type of recall and involve situations where there is a reasonable probability that use of the product will cause serious injury or death. |
For additional information on this product recall, see the Carl Zeiss Meditec press release at: http://www.fda.gov/cdrh/recalls/recall-060304-pressrelease.html Also, for additional information on this product recall, see the FDA Enforcement Report located at: http://www.fda.gov/bbs/topics/enforce/2004/ENF00856.html. Scroll down to "Recalls and Field Corrections: Devices – Class I". |
Updated July 30, 2004
CDRH Home Page | CDRH A-Z Index | Contact CDRH | Accessibility | Disclaimer
FDA Home Page | Search FDA Site | FDA A-Z Index | Contact FDA | HHS Home Page
Center for Devices and Radiological Health / CDRH