NEMI Method Summary

Method Summary Information

Method Number: 9056A Media: VARIOUS
Revision: Revision 1, November 2000
Method Source: U.S. EPA Office of Solid Waste
Subcategory: INORGANIC Analytes in this method
Official Name: Determination Of Inorgainic Anions By Ion Chromatography
Descriptive Name Anion Chromatography Method
Source Info: U.S. EPA Office of Solid Waste (OSW) Methods Team
Ariel Rios Bldg. (5307W)
1200 Pennsylvania Ave. NW
Washington, DC 20460
Phone: 703-308-8855
Fax: 703-308-0511
Citation: SW-846 Update IVB
51305 Byte file
Brief Method
Summary:
A small volume (2-3 mL) of sample is injected into an ion chromatograph to flush and fill a constant volume sample loop. The sample is then injected into a stream of carbonate-bicarbonate eluent of the same strength as the sample. The samples are passed through three ion exchange columns (with different functionalities), and compounds are identified by retention time, and quantitated by peak height or area (by comparison with a standard calibration curve).
Scope And
Application:
This method is applicable to the determination of chloride, fluoride, bromide, nitrate, phosphate, and sulfate in the collection solutions from the bomb combustion of solid waste samples, as well as all water samples.
Applicable
Conc Range:
0.2 - 100mg/L. Ranges differ for each analyte. This range is taken from the lowest and highest limit of the analytes listed in the method.
Interferences:
QC Requirements: Mid-range calibration standard every 10 injections; sample duplicate (1 in every 10 samples), MS
Sample Handling: Collect the samples using a sampling plan that addresses the considerations discussed in Chapter 9 of this manual. Preserve the samples by refrigeration at 4 degrees C. Analyze the samples as soon as possible after collection.
Max Holding Time:
ASAP
 
Relative Cost: $51 to $200