NEMI Method Summary

Method Summary Information

Method Number: D1293A Media: WATER
Revision: 1995
Method Source: ASTM International
Subcategory: PHYSICAL Analytes in this method
Official Name: Standard Test Methods for pH of Water
Descriptive Name pH: Method A: Precise Laboratory Measurement of pH
Source Info: ASTM International
100 Barr Harbor Drive West
Conshohocken, PA 19428
Telephone: (610) 832-9500
Fax: (610) 832-9555
Email: service@astm.org
Citation: ASTM Volume 11.01
 
Brief Method
Summary:
The pH meter and associated electrodes are standardized against two reference buffer solutions that closely bracket the anticipated sample pH. The sample measurement is made under strictly controlled conditions and prescribed techniques.
Scope And
Application:
This test method covers the precise measurement of pH in water under strictly controlled laboratory conditions.
Applicable
Conc Range:
1.0 - 12.0 pH
Interferences: The glass electrode reliably measures pH in nearly all aqueous solutions and in general is not subject to solution interference from color, turbidity, colloidal matter, oxidants, or reductants. The glass electrode reliably measures pH in nearly all aqueous solutions and in general is not subject to solution interference from color, turbidity, colloidal matter, oxidants, or reductants. The reference electrode may be subject to interferences and should be chosen to conform to all requirements of the Method.
QC Requirements:
Sample Handling: Collect samples in accordance with Practice D 1066, Specification D 1192, or Practices D 3370, whichever is applicable.
Max Holding Time:
 
Relative Cost: Less than $50
URL to obtain ordering information for this method:
http://www.astm.org/cgi-bin/SoftCart.exe/DATABASE.CART/PAGES/D1293.htm?E+mystore
4748
Page 1