NEMI Method Summary

Method Summary Information

Method Number: 8100 Media: VARIOUS
Revision: Revision 0, September 1986
Method Source: U.S. EPA Office of Solid Waste
Subcategory: ORGANIC Analytes in this method
Official Name: Polynuclear Aromatic Hydrocarbons
Descriptive Name Polynuclear Aromatic Hydrocarbons in Water and Solid Samples by GC-FID
Source Info: U.S. EPA Office of Solid Waste (OSW) Methods Team
Ariel Rios Bldg. (5307W)
1200 Pennsylvania Ave. NW
Washington, DC 20460
Phone: 703-308-8855
Fax: 703-308-0511
Citation: SW-846 On-line
212312 Byte file
Brief Method
Summary:
Samples are extracted and cleaned up (according to sample matrix) and then are directly injected into a gas chromatograph using the solvent flush technique and are detected by a flame ionization detector.
Scope And
Application:
This method determines the concentration of polynuclear aromatic hydrocarbons (PAHs) in water and solid samples using gas chromatography and flame ionization detection.
Applicable
Conc Range:
Not Available.
Interferences: (A) Method interferences: Contaminants in solvents, reagents, glassware, and other sample processing hardware may lead to elevated baselines and/or other interferences in the gas chromatograms. Routine analysis of reagent blanks should be performed as a check for method interferences. Proper selection of reagents and purification of solvents by distillation in all-glass systems may be required.
(B) Matrix interferences: Unique samples may require additional cleanup.
(C) Refer to Methods 3500, 3600, and 8000.
QC Requirements: Refer to Chapter One and Methods 8000, 3500, and 3600 for specific QC procedures. Procedures include: validation of extraction and clean up techniques, analysis of quality control check samples, method blanks, matrix spikes, and surrogates. In addition, refer to Method 8000, Section 8.6 for procedures to check the GC system operation.
Sample Handling: See SW-846 Chapter Four for information. See the introductory material to this chapter, Organic Analytes, Section 4.1. Extracts must be stored under refrigeration.
Max Holding Time:
Aqueous samples - 7 days (sample), 40 days (extract). Solid samples - 14 days (sample), 40 (extract). Also, see SW-846 Chapter Four for information.
Sample Prep Met Methods 3510, 3520, 3540, or 3550
Relative Cost: $201 to $400
5256