NEMI Method Summary

Method Summary Information

Method Number: 208.1 Media: WATER
Revision: Issued 1974
Method Source: U.S. EPA National Exposure Research Laboratory (NERL) [formerly EMSL]
Analytes in this method
Subcategory: INORGANIC
Official Name: Barium (Atomic Absorption, Direct Aspiration)
Descriptive Name Barium by Flame AA
Source Info: U.S.EPA National Exposure Research Laboratory (NERL)
Microbiological and Chemical Exposure Assessment Research Division (MCEARD)
[formerly the Environmental Monitoring Systems Laboratory (EMSL), Cincinnati, OH]
26 West Martin Luther King Drive
Cincinnati, Ohio 45268-0001
Fax: 513-569-7757
email: DWmethods.help@epa.gov
Citation: Methods for the Chemical Analysis of Water and Wastes (MCAWW) (EPA/600/4-79/020)
12929 Byte file
Brief Method
Summary:
A sample prepared according to EPA Method 200.0 is analyzed using a flame atomic absorption spectrometer. Specifications for instrument operation and preparation of standards are provided in the method.
Scope And
Application:
This method determines barium by flame atomic absorption spectrometry. Sample pretreatment according to EPA Method 200.0 determines the applicability.
Applicable
Conc Range:
1,000 - 20,000 ug/L (optimal range); 400 ug/L (sensitivity).
Interferences: (A) Ionization: Barium is easily ionized by the oxide-acetylene flame. Add potassium (1000 mg/L) to samples and standards to control this effect.
(B) Acetylene-Air: If acetylene-air is used, phosphate, silicon, and aluminum will severely suppress the barium absorbance. This may be overcome by adding 2000 mg/L of lanthanum.
For additional interferences see EPA Method 200.0.
QC Requirements: QC requirements for drinking water procedures are provided in method 200.0.
Sample Handling: Sample handling and preservation procedures are provided in EPA Method 200.0.
Max Holding Time:
6 months (MCAWW, Table 1).
Sample Prep Met 200.0
Relative Cost: Less than $50
5275