NEMI Method Summary

Method Summary Information

Method Number: 218.2 Media: WATER
Revision: Issued 1978
Method Source: U.S. EPA National Exposure Research Laboratory (NERL) [formerly EMSL]
Analytes in this method
Subcategory: INORGANIC
Official Name: Chromium (Atomic Absorption, Furnace Technique)
Descriptive Name Chromium by Graphite Furnace AA
Source Info: U.S.EPA National Exposure Research Laboratory (NERL)
Microbiological and Chemical Exposure Assessment Research Division (MCEARD)
[formerly the Environmental Monitoring Systems Laboratory (EMSL), Cincinnati, OH]
26 West Martin Luther King Drive
Cincinnati, Ohio 45268-0001
Fax: 513-569-7757
email: DWmethods.help@epa.gov
Citation: Methods for the Chemical Analysis of Water and Wastes (MCAWW) (EPA/600/4-79/020)
13070 Byte file
Brief Method
Summary:
A sample prepared according to EPA Method 200.0 is analyzed using a graphite furnace atomic absorption spectrometer. Specifications for instrument operation and preparation of standards are provided in the method.
Scope And
Application:
This method determines chromium by graphite furnace atomic absorption spectrometry. Sample pretreatment according to EPA Method 200.0 determines the applicability.
Applicable
Conc Range:
5-100 ug/L (optimal concentration).
Interferences: (A) Dissolved solids: Use background correction to prevent error from high dissolved solid levels.
(B) Contamination: Pipette tips may contaminate samples.
(C) Cyanide: Nitrogen gas should be avoided because of possible cyanide band interference.
For additional interferences see EPA Method 200.0.
QC Requirements: QC requirements for drinking water procedures are provided in method 200.0
Sample Handling: Sample handling and preservation procedures are provided in EPA Method 200.0.
Max Holding Time:
6 months (MCAWW, Table 1).
Sample Prep Met 200.0
Relative Cost: $51 to $200
5283